Test-retest repeatability of MBSS-T1: a model-based, subject-specific self-supervised motion correction method for robust cardiac T1 mapping.
Adi Zehavi Lenz, Eyal Hanania, Ilya Volovik, Daphna Link-Sourani, Moti Freiman
Faculty of Biomedical Engineering, Technion Israel Institute of Technology, Technion City, Haifa, 3200003, Israel. · Faculty of Electrical and Computer Engineering, Technion Israel Institute of Technology, Technion City, Haifa, Haifa District, 3200003, Israel.
Key points
From the paper's conclusion- This study evaluated the test–retest repeatability of MBSS-T1, a subject-specific, model-based, self-supervised deep-learning method for motion-robust myocardial T1 mapping.
Approach.
Abstract
Reliable myocardial T1 mapping is essential for quantitative cardiac tissue characterization but remains sensitive to respiratory motion, particularly during free-breathing acquisitions. This study evaluated the test-retest repeatability of MBSS-T1, a subject-specific, model-based, self-supervised deep-learning method for motion-robust myocardial T1 mapping. Approach. Seventeen healthy male volunteers aged 20-36 years underwent 3 T cardiac MRI using a 5(3)3 MOLLI acquisition under breath-hold (BH) and free-breathing (FB) conditions. Same-day repeat scans were acquired after a 20 min interval that included removal from and repositioning in the scanner. T1 maps were reconstructed using (1) MBSS-T1, (2) the contrast-agnostic deep-learning registration method SynthMorph, (3) scanner-inline motion correction, and (4) no motion correction. Repeatability was evaluated using intraclass correlation coefficients (ICC), relative Bland-Altman analysis, within-subject coefficients of variation (CVw), relative 95% repeatability coefficients (RC95), and variance-based bootstrap inference with Holm correction. Main results. All four methods included 17 complete participant-level test-retest pairs under each breathing condition. Under BH conditions, the global ICC point estimates were 0.77, 0.64, 0.65, and 0.60 for MBSS-T1, SynthMorph, scanner-inline correction, and no correction, respectively. Under FB conditions, the corresponding values were 0.41, 0.32, 0.27, and 0.16. MBSS-T1 also had the lowest descriptive CVw under both conditions. However, none of the paired comparisons of absolute test-retest differences was significant after Holm correction (all adjusted p ≥ 0.623). The smallest Holm-adjusted variance-ratio result was obtained for MBSS-T1 versus no correction under FB conditions (variance ratio 0.43, 95% CI 0.30-0.68; adjusted p = 0.058), which did not meet the prespecified significance threshold. Significance. MBSS-T1 yielded favourable descriptive repeatability estimates, particularly under FB conditions, but no pairwise comparison remained statistically significant after correction for multiple comparisons. These results characterize same-day, same-scanner measurement precision rather than absolute T1 accuracy and require confirmation in larger and more clinically diverse cohorts.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Native myocardial T1 mapping using rapid 1.5T T1FLASH and T1 MOLLI during breath-hold and free-breathing.BMC cardiovascular disorders · 2026
- Training-free motion correction for cardiac T1 mapping using DINOv3 features.Magnetic resonance imaging · 2026
- High-Receiver-Bandwidth MRI Near Metal Implants with Motion-Consistent Rep2Rep DenoisingarXiv · 2026
- Self-supervised reconstruction framework via motion- and physics-informed learning for four-dimensional magnetic resonance fingerprinting.Medical image analysis · 2026
- Improved Myelin Water Fraction Estimation and Uncertainty Mapping in Multicomponent T2 Relaxometry via NNLS Bootstrap AggregationarXiv · 2026
- Training-Free Motion-Corrected 3D Brain MRI Reconstruction via Event-Guided ReferencesarXiv · 2026