Untangling the causes of elongation artifacts in dual flat-panel PET: limited angular sampling versus parallax error.
Boris Vervenne, Jens Maebe, Maya Abi Akl, Christian Vanhove, Stefaan Vandenberghe
Department of Electronics and Information Systems, Ghent University, Corneel Heymanslaan 10, entrance 37, Ghent, Flanders, 9000, Belgium. · Department of Electronics and Information Systems, Ghent University, Corneel Heymanslaan 10, entrance 37, Gent, Flanders, 9000, Belgium.
Key points
From the paper's conclusion- Elongations due to limited angular coverage in the flat-panel system grew with object size and dominated for features over 1-2 cm.
Abstract
OBJECTIVE: Dual flat-panel PET systems provide easy patient access and high sensitivity when the panels are close together. Their images are however susceptible to elongation artifacts due to incomplete angular sampling and parallax error. These effects are reduced by respectively including depth-of-interaction (DOI) and time-of-flight (TOF) information. To guide system design, a framework is needed that predicts which effect dominates for a given object size and field-of-view (FOV) location. This study investigates the impact of TOF and DOI on elongation artifacts for a variety of phantoms in a dual flat-panel system, benchmarked against a cylindrical reference. APPROACH: A large dual flat-panel system and cylindrical system with the same axial FOV were simulated. The TOF and DOI resolution were varied independently and iterative MLEM reconstruction was performed. Data were generated for point sources and spheres spanning 1-50 mm radius at multiple FOV positions. Point-source blur was quantified with full width at half maximum measurements, and sphere elongation using an activity-spillover metric. Findings were validated using the more clinically relevant NEMA IQ and XCAT phantoms with embedded lesions. MAIN RESULTS: DOI uncertainty resulted in elongations independently of object size. In the cylindrical system, these only occurred at radial offsets, but in the dual flat-panel system, parallax-blur was also present at the center of the FOV. Elongations due to limited angular coverage in the flat-panel system grew with object size and dominated for features over 1-2 cm. The extent of these elongations increased towards the FOV edges and was substantially reduced by including TOF information. With adequate TOF and DOI performance, the flat-panel system matched cylindrical image quality. SIGNIFICANCE: The framework predicts the dominant artifact mechanism as a function of object size, position, and system geometry. The insights guide detector and system design for dual flat-panel PET and corresponding artifact correction methods.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of depth-of-interaction and spatially-variant point-spread-function model on the performance of ultra-high resolution brain PET.Physics in medicine and biology · 2026
- Multi-GPU parallel framework for DOI-enabled long axial field-of-view PET image reconstruction.Physics in medicine and biology · 2026
- Beam-width-corrected DOI resolution measurement of dual-ended readout PET detectors.Physics in medicine and biology · 2026
- Single-Shot and Rotational Stitching Methods for Long-Length FPD Imaging.Radiologic technology · 2026
- PhysiCal-PET: Physics-driven neural networks for the rapid position Calibration of PET ScannersarXiv · 2026
- PSF-Informed Reconstruction: An Enhanced 2D Directional Filtered Backprojection Method for Cone-Beam Rotary Computed LaminographyarXiv · 2026