Detection and correction of rigid patient motion in SPECT with exponential data consistency conditions.
My Hoang Hoa Bui, Antoine Robert, Jean-Noël Badel, Ane Etxebeste, Simon Rit
CREATIS, 21 avenue Jean Capelle, Villeurbanne, 69621, France. · CREATIS, 21 avenue Jean Capelle, Villeurbanne, Auvergne-Rhône-Alpes, 69621, France.
Key points
From the paper's conclusion- Specifically, the method achieved a relative activity recovery coefficient (rARC) of around 101% relative to images reconstructed from motion-free projections, compared with an rARC of 87% without correction for the studied liver patient's rigid motion occurring at the beginning of the acquisition. 

Significant A fully data-driven method with eDCCs was able to estimate and correct for a r
Abstract
Objective Rigid patient motion during single photon emission computed tomography (SPECT) can introduce artifacts in the reconstructed images. Patient translation can be estimated using various data-driven methods, e.g., those based on exponential data consistency conditions (eDCCs). eDCCs are mathematical equations that characterize the small redundancy between pairs of exponential projections derived from the application of the attenuated Radon transform. However, rigid motion typically involves both translational and rotational components. This study aimed to develop and evaluate an eDCCs-based approach for detecting and estimating any rigid patient motion during a SPECT acquisition. Approach We calculated exponential projections from SPECT measurements, assuming that the photon emission was confined to a known convex region K with constant attenuation, and evaluated their pairwise consistency with eDCCs. The rigid motion was characterized by six parameters, three translations and three rotations, and the index indicating when the motion occurred. The six parameters and the motion index were estimated by minimizing the inconsistency between pairs of exponential projections after compensating for the estimated rigid motion in the projection domain. The method was evaluated on several datasets of a simulated liver patient and on experimentally acquired data of a NEMA IEC phantom. Main results The proposed method accurately estimated the ground truth motion parameters when the SPECT projections were ideally modeled by the attenuated Radon transform, but was affected by degrading physical effects, mainly by the system spatial resolution (intrinsic and collimator spatial resolution) and the scatter. Despite these physical effects, eDCCs recovered the source center position within a two-pixel offset and substantially improved reconstructed image quality. Specifically, the method achieved a relative activity recovery coefficient (rARC) of around 101% relative to images reconstructed from motion-free projections, compared with an rARC of 87% without correction for the studied liver patient's rigid motion occurring at the beginning of the acquisition. Significant A fully data-driven method with eDCCs was able to estimate and correct for a rigid patient motion during a SPECT acquisition.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Applying exponential data consistency conditions to non-homogeneous attenuation.Physics in medicine and biology · 2026
- Rigid Motion Estimation Using Accelerated Coordinate Descent (REACT) for MR Imaging.Magnetic resonance in medicine · 2026
- Data-driven motion compensation in HR-pQCT.Biomedical physics & engineering express · 2026
- Quantification of patient motion in FDG-PET/CT brain imagingmedRxiv · 2026
- Radionuclide Dependence of Distance-Dependent Spatial Resolution and Resolution Recovery in SPECT: A SIMIND Monte Carlo Simulation Study.Biomedical physics & engineering express · 2026
- Kinematic Total-Body PET Video Reconstruction via Hierarchical List-Mode EM and Validation-Adaptive Space-Time RegularisationarXiv · 2026