Rigid Motion Estimation Using Accelerated Coordinate Descent (REACT) for MR Imaging.
Kwang Eun Jang, Dwight G Nishimura
Independent Researcher, Busan, Republic of Korea. · Magnetic Resonance Systems Research Lab (MRSRL), Department of Electrical Engineering, Stanford University, Stanford, California, USA.
Key points
From the paper's conclusion- This study demonstrates the feasibility of coordinate descent for autofocus motion correction in MR imaging.
Abstract
PURPOSE: To develop a computationally viable autofocus method for estimating 3D rigid motion in MR imaging. THEORY AND METHODS: The proposed method, REACT, assumes a piecewise-constant motion trajectory and estimates the rigid motion parameters of individual temporal segments by optimizing an image-quality metric. Coordinate descent is adopted to decompose the high-dimensional optimization problem into a series of subproblems, each updating the motion parameters of a single temporal segment. The cost function of each subproblem is assumed to be approximately locally convex under suitable acquisition conditions. Each subproblem is then solved using a derivative-free solver, thereby avoiding an exhaustive grid search. Numerical simulations investigated the local convexity assumption and data acquisition requirements. REACT was evaluated for respiratory motion correction on in vivo free-breathing coronary MR angiography datasets. Coronary artery sharpness was quantified using unbounded image edge profile acutance (u-IEPA). RESULTS: In numerical simulations, the objective surfaces of the subproblems were approximately locally convex when the current motion estimate was sufficiently close to the desired solution, and REACT required the data collected within each temporal segment to be sufficiently distributed across k -space. In the in vivo study, REACT yielded higher u-IEPA for both the left anterior descending artery (LAD) and the right coronary artery than did a conventional translational motion-estimation method using image-based navigators. REACT also yielded higher u-IEPA for the LAD than did a conventional autofocus nonrigid motion correction method. CONCLUSION: This study demonstrates the feasibility of coordinate descent for autofocus motion correction in MR imaging.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- DoReMe: Dual optimization of Reconstruction and Motion estimation for self-navigated free-breathing abdominal MRI using Gaussian SplattingarXiv · 2026
- Non-Rigid Motion-Compensated MR Multitasking for Free-Breathing Low-Dose Dynamic Contrast-Enhanced MRI in the Abdomen.Magnetic resonance in medicine · 2026
- An expert-supervised numerical framework for 3D coronary artery reconstruction and blood flow velocity estimation from biplane angiographyRussian Journal of Numerical Analysis and Mathematical Modelling · 2026
- Detection and correction of rigid patient motion in SPECT with exponential data consistency conditions.Physics in medicine and biology · 2026
- Iterative Navigator-Based Retrospective Motion Correction for Multi-Shot Brain MR Elastography and Evaluation of Single-Shot Post-Reconstruction Correction.Magnetic resonance in medicine · 2026
- Motion Mitigation Techniques for Abdominal and Cardiac MR ImagingJournal of Magnetic Resonance Imaging · 2025