Effects of count rate performance on image quality in an LAFOV PET/CT system.
Nanna Overbeck, Zuzanna Czeluśniak, Søren Holm, Julie Verne Henriksen, Flemming L Andersen, Thomas Lund Andersen
Department of Clinical Physiology and Nuclear Medicine, Copenhagen University Hospital -Rigshospitalet, Blegdamsvej 9, Copenhagen, 2100, Denmark.
Key points
From the paper's conclusion- This study investigated the relationship between body size, administered activity (18F), and image quality using two patient-like phantoms representing normal (78.5 kg) and obese (194.3 kg) body profiles.
Abstract
The development of Long Axial Field-of-View (LAFOV) positron emission tomography (PET) combined with computed tomography (CT) warrants reconsideration of the administered activity required for clinical imaging. This study investigated the relationship between body size, administered activity (18F), and image quality using two patient-like phantoms representing normal (78.5 kg) and obese (194.3 kg) body profiles. The phantoms were scanned in selected five-minute periods within a total duration of 27 and 21 hours, respectively, following the decay of the activity. Analysis of the trues, scatter, and randoms demonstrated a close to linear propagation of the trues with activity up to a point where saturation was reached. The normal body profile phantom achieved higher signal-to-noise ratios (SNRs), reaching system saturation at a lower activity (680 MBq) than the large phantom, representing the obese body profile (970 MBq). The SNR2 showed a nonlinear correlation with the noise equivalent count rate (NECR), whereas SNR2 had a linear correlation with the trues. The study demonstrated the count rate capabilities up to and beyond the saturation limit. Furthermore, we illustrated the possibilities of scanning patients with ultralow dose administration in the range of 0.03-0.07 MBq/kg body weight. However, achieving robust clinical diagnostic quality - particularly for low-contrast lesions - at these activity levels requires further clinical investigation. Utilizing these ultralow activity concentrations requires extending the scan duration to compensate for image quality, especially when examining obese patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- EARL-based quantitative evaluation of administered activity reduction for [68Ga]Ga-DOTATOC PET/CT on a 32-cm axial FOV digital BGO system: Phantom analysis and retrospective list-mode clinical assessment.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Simulation of low-dose PET imaging protocols for assessment of pancreatic beta-cell mass in pediatric type 1 diabetesmedRxiv · 2026
- Robustness of Long Axial Field-of-View PET to Defective Detector Blocks: Impact on Quantitative AccuracymedRxiv · 2026
- Optimization of patient dose and image quality for positron emission tomography-computed tomography.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Intraindividual Comparison of Radiation Dose and Image Quality: Photon-counting versus Energy-integrating Detector CT in Obesity.Radiology · 2026
- Automated long axial field of view PET image processing and kinetic modelling with the TurBO toolboxEuropean Journal of Nuclear Medicine and Molecular Imaging · 2026