An automated geometric modeling framework in GATE for the design and optimization of high-sensitivity converging-beam SPECT collimators.
Fenghua Weng, Gaoyu Chen, Yunlong Zan, Greta S P Mok, Youngho Seo, Qiu Huang
Macao Polytechnic University, Rua de Luís Gonzaga Gomes, Macao, China, Macau, 999078, China. · Vplus Technology Ltd., 2nd Unit, Building F, Feimalv, Minhang District, Shanghai, China, ShangHai, ShangHai, 200240, China.
Key points
From the paper's conclusion- The results demonstrate that the proposed CBC design offers a significant sensitivity advantage, making it highly suitable for high-performance, small-volume clinical applications such as brain and cardiac molecular imaging. \end{abstract}.
Abstract
Objective:The trade-off between detection sensitivity and spatial resolution is a fundamental challenge in designing organ-dedicated SPECT collimators. While converging-hole geometries offer a solution, their optimization is often hindered by the lack of flexible computational tools capable of modeling large-scale, non-parallel hole arrays. This study aims to develop an automated geometric modeling framework to facilitate the design and evaluation of complex converging- and diverging-hole collimators within standard Monte Carlo environments. Approach:We developed a specialized modeling framework by implementing custom C++ classes and a vector-based alignment algorithm within GATE. This platform enables automated, orientation-consistent construction of large-scale converging arrays not natively supported by standard implementations. A high-sensitivity pure cone-beam collimator (CBC) was designed using this framework. The evaluation used hot-rod, disc, and Jaszczak phantoms for physical characterization, while XCAT and dedicated brain models were employed for clinical tasks, including cardiac, brain perfusion, and DaTscan SPECT simulations. Main results:The CBC achieved a nearly fourfold sensitivity increase compared to a conventional low-energy high-resolution (LEHR) parallel-hole collimator at a 20 cm radius of rotation, while maintaining comparable spatial resolution. Despite a 52.3% FOV reduction, the CBC yielded a 2.2-fold noise reduction (CV: 11.7% vs. 25.9%) and mitigated PVE via geometric magnification. XCAT and brain phantom simulations confirmed enhanced anatomical definition and contrast recovery in cardiac, perfusion, and DaTscan tasks. Significance:This work provides an efficient computational tool for rapid design space exploration of advanced collimator geometries. The results demonstrate that the proposed CBC design offers a significant sensitivity advantage, making it highly suitable for high-performance, small-volume clinical applications such as brain and cardiac molecular imaging. \end{abstract}.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- OpenTOPAS-SPECT: an extension for full-physics clinical and preclinical SPECT simulation.Physics in medicine and biology · 2026
- Geometry-conditioned differentiable surrogate with adjoint Monte Carlo sampling for scintillator detector co-designarXiv · 2026
- Eliminating Registration Bias in Synthetic CT Generation using a physics-based simulation framework for pelvic anatomy.Physics in medicine and biology · 2026
- Gradient-Based Trajectory Optimisation over Continuous Poses for Sparse-View Cone-Beam CTarXiv · 2026
- Confidence-guided cryo-EM map optimisation with LocScale-2.0Nature Communications · 2026
- Multi-GPU parallel framework for DOI-enabled long axial field-of-view PET image reconstruction.Physics in medicine and biology · 2026