The comparison and application of the measurement method for radon progeny activity size distribution based on SDB and MLP.
Duanyang Ma, Jinmin Yang, 康 彭, Lei Zhang
State Key Laboratory of Chemistry for NBC Hazards Protection, No. 37 Yuan, Central South Street, Changping District, Beijing, 102205, China. · State Key Laboratory of Nuclear Physics and Technology, Peking University, No. 5 Summer Palace Road, Haidian District, Beijing, 100871, China.
Key points
From the paper's conclusion- And the activity size distribution of the attached radon progeny in the outdoor environment all exhibits a bimodal distribution with the average diameter of the nucleation mode is 22 nm and that of the accumulation mode is 229 nm, and only a small portion of those show a unimodal distribution in the indoor environment, while the average diameters of the nucleation mode and accumulation mode separa
Abstract
To achieve the accurate measurement of the activity size distribution of radon progeny and identify the distribution mode automatically, a measurement method based on the screen diffusion battery (SDB) and the Multilayer Perceptron (MLP) machine learning algorithm was developed in this study. Comparison experiments and field measurements were also carried out. Results show that the MLP algorithm presents superior performance in accurately identifying the unimodal or bimodal distribution and avoiding unambiguous solutions compared with the traditional ex-pectation maximization (EM) and Twomey inversion algorithms. Meanwhile, the SDB+MLP method proposed in this study has principle-based advantages over the conventional impactor method and SMPS indirect measurement method for solving parameters and the fraction of small-sized nucleation mode radon progeny. Field measurement results show that the SDB+MLP system can be applied to the activity size distribution measurement of the attached radon progeny. And the activity size distribution of the attached radon progeny in the outdoor environment all exhibits a bimodal distribution with the average diameter of the nucleation mode is 22 nm and that of the accumulation mode is 229 nm, and only a small portion of those show a unimodal distribution in the indoor environment, while the average diameters of the nucleation mode and accumulation mode separately are 25 nm and 209 nm. For the average fraction of the accumulation mode, a little difference of 66% and 69% for indoors and outdoors is shown, and the range is from 52% to 84% for outdoors and 49% to 87% for indoors. 
.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Novel data-processing method for determining the activity size distribution of unattached radon progeny with multi-layer perceptron machine learning algorithm.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Measurement of activity size distribution of radon and thoron progeny using DRPS/DTPS based cascade impactor.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Machine Learning-Enhanced NDIR Methane Sensing Solution for Robust Outdoor Continuous Monitoring ApplicationsSensors · 2025
- Factors associated with radon occurrence in drinking water in Montenegro: a statistical and machine learning approach.Environmental geochemistry and health · 2026
- Robust Prediction-Reliability Assessment in Biomedical Sensing and Classification Under Data Scarcity Using Conformal Prediction and Statistical Randomization.Sensors (Basel, Switzerland) · 2026
- Machine learning models for predicting plastic bronchitis in hospitalized children with Mycoplasma pneumoniae pneumonia undergoing clinically indicated bronchoscopy: a single-center retrospective study.Frontiers in pediatrics · 2026