Journal of Physics D: Applied Physics

Self-biased magnetoelectric Ni/LiNbO₃/Ni composites: from material characterisation to a battery-free implantable node

Hakeim Talleb, Manar Touati, Samir Beldi, Yunlin Zheng, Aurélie Gensbittel, Erwan Dandeu, Spencerh Bidouba, Tianwen Huang,

Key points

From the paper's conclusion
  • This power is shown, by circuit simulation on the measured source and by a measured-power budget, to run a complete battery-free node: the ME link supplies a pressure sensor, its analogue-to-digital converter and a microcontroller, a 13.56 MHz backscatter channel returns the data, and the harvested power lets a semi-active uplink clear the range limit of passive backscatter at 10 cm.

Abstract

Self-biased magnetoelectric (ME) composites, which convert a magnetic field into electricity with no static bias, are attractive for powering battery-free, deeply implanted microsystems through tissue without heating it, while returning their data. This paper develops a self-biased Ni/LiNbO₃/Ni composite in which sputter-deposited nickel acts both as the magnetostrictive phase and as the electrodes; an intrinsic residual stress (≈ 50 MPa, confirmed by temperature-dependent VSM) supplies the bias and removes the static field, permanent magnets and epoxy bonding of conventional ME laminates. A unified equivalent circuit, a Mason model reduced near resonance to a Butterworth–Van Dyke network closed by a magneto-mechanical gyrator and a piezoelectric transformer, all elements extracted in closed form from impedance and VSM data, predicts the deliverable power within a factor of two across four composites over two decades of output, and is ready for end-to-end co-simulation in SPICE (Simulation Program with Integrated Circuit Emphasis). Guided by a fill-factor and skin-depth design study, an optimised sample raises the output from ≈ 1 V / 16 µW to 8.5 V / 0.33 mW self-biased (1.5 mW under a small bias). This power is shown, by circuit simulation on the measured source and by a measured-power budget, to run a complete battery-free node: the ME link supplies a pressure sensor, its analogue-to-digital converter and a microcontroller, a 13.56 MHz backscatter channel returns the data, and the harvested power lets a semi-active uplink clear the range limit of passive backscatter at 10 cm. A dosimetric and thermal analysis shows the low operating frequency to be a primary safety enabler (induced field and SAR scale as f and f²): the ME link deposits 50 to 130 times less specific absorption rate than an iso-footprint 13.56 MHz inductive link operated at the field its own standard permits.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Self-biased magnetoelectric Ni/LiNbO₃/Ni composites: from material characterisation to a battery-free implantable node — Science Skim