Bioanalysis

A novel method for the quantification of vanzacaftor, deutivacaftor, lumacaftor, elexacaftor, tezacaftor, ivacaftor, and selected metabolites in human plasma using liquid chromatography-tandem mass spectrometry.

Sarah Thomas, David Joyce, Natalia Popowicz, Siobhain Mulrennan, Sam Salman

School of Health and Clinical Sciences, Faculty of Health and Medical Sciences, University of Western Australia, Crawley, Western Australia, Australia. · Clinical Biochemistry Pharmacology and Toxicology, PathWest Laboratory Medicine, Nedlands, Western Australia, Australia.

Key points

From the paper's conclusion
  • This validated LC-MS/MS method is robust, sensitive, and suitable for routine clinical application. The short runtime and minimal sample volume support its use in high-throughput settings.

Abstract

BACKGROUND: Cystic fibrosis transmembrane conductance regulator modulators (CFTRm) have revolutionized the care of patients with cystic fibrosis. There are several clinical justifications for CFTRm therapeutic drug monitoring. To facilitate the implementation of this service, we developed a liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for quantifying CFTRm concentrations in human plasma. METHODS: A high-throughput LC-MS/MS assay was developed for the simultaneous quantification of current and emerging CFTRm, including deutivacaftor, vanzacaftor, ivacaftor, lumacaftor, tezacaftor and elexacaftor, along with major metabolites (hydroxymethyl-ivacaftor, ivacaftor-carboxylate, tezacaftor-M1, and elexacaftor-M23). Sample preparation involved protein precipitation using 5 μL plasma, ensuring efficiency and minimal sample requirement. Chromatographic separation was achieved using a C18 column. Detection was performed via positive electrospray ionization in multiple-reaction-monitoring mode, with stable isotope-labeled internal standards. The instrument runtime was 3.5 minutes, supporting rapid analysis for a high-throughput laboratory. RESULTS: The assay demonstrated a linear range of 0.01 to 10 mg/L for all analytes. Method validation demonstrated high accuracy and precision. Stability studies indicated that all analytes remained stable in plasma under a range of tested conditions. CONCLUSIONS: This validated LC-MS/MS method is robust, sensitive, and suitable for routine clinical application. The short runtime and minimal sample volume support its use in high-throughput settings.

Read the paper