Intraoperative electrically evoked stapedial reflex thresholds (eSRT) for cochlear implant programming: a postoperative validation study.
Eric Cunningham, Isabelle J Chau, Rivka Bornstein, Regina Albinus, Pal Sudhirkumar Shah, Katrina Stidham
School of Medicine, New York Medical College, Valhalla, NY, USA. · Department of Otolaryngology-Head & Neck Surgery, Westchester Medical Center, Valhalla, NY, USA.
Key points
From the paper's conclusion- Intraoperative eSRT was feasible and yielded high response rates, particularly at apical and middle electrodes. However, intraoperative thresholds did not correlate with early postoperative values and cannot currently predict early postoperative eSRTs. These findings support the utility of intraoperative eSRT as a qualitative intraoperative tool, but further refinement is needed to improve its predictive value for postoperative cochlear implant programming.
Abstract
OBJECTIVE: To determine if intraoperative electrically evoked stapedial reflex thresholds (eSRT) correlate with postoperative eSRT and can predict programming levels in cochlear implant patients. DESIGN: Prospective cohort study conducted at a tertiary hospital. Seventy-seven patients who underwent cochlear implant surgery were included. Intraoperative and postoperative eSRT testing was performed at electrodes E1, E6, E12, E17, and E22, with postoperative testing conducted at a median of 4 weeks. RESULTS: Eighty-four cochlear implants in 77 patients met inclusion criteria, with intraoperative eSRT attempted in 72 ears and postoperative testing in 50. Intraoperative response rates were significantly higher than postoperative rates overall (83.1% vs. 70.9%, p = 0.0005), with the highest rates at middle and apical electrodes. Median intraoperative thresholds were significantly higher than postoperative thresholds (21.2 vs. 15.6, p < 0.0001), but no significant correlations were observed between the two time points. CONCLUSION: Intraoperative eSRT was feasible and yielded high response rates, particularly at apical and middle electrodes. However, intraoperative thresholds did not correlate with early postoperative values and cannot currently predict early postoperative eSRTs. These findings support the utility of intraoperative eSRT as a qualitative intraoperative tool, but further refinement is needed to improve its predictive value for postoperative cochlear implant programming.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optimizing Cochlear Implant Programming in Single‐Sided Deafness and Asymmetric Hearing Loss: A Randomized Comparison of ESRT and Behavioral MethodsLaryngoscope Investigative Otolaryngology · 2026
- Electromyography of the stapedius muscle via a retrofacial approach and electrically evoked stapedius reflex during cochlear implant surgery: a prospective bicentric studyScientific Reports · 2026
- Objective measurements of subperiosteal pocket migration for cochlear implant: long-term results from a prospective study.Cochlear implants international · 2026
- Simultaneous Intraoperative eCAP and eABR Recordings During Cochlear Implant Surgery Reveal Complementary Neural Response Characteristics.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026
- Stapedius reflex measurements in open sound-field in children using MED-EL cochlear implant systems.International journal of audiology · 2026
- Remote Check in Cochlear Implant Follow-Up: An Exploratory Comparison with Conventional Audiological Measurements.Audiology research · 2026