Cochlear implants international

Intraoperative electrically evoked stapedial reflex thresholds (eSRT) for cochlear implant programming: a postoperative validation study.

Eric Cunningham, Isabelle J Chau, Rivka Bornstein, Regina Albinus, Pal Sudhirkumar Shah, Katrina Stidham

School of Medicine, New York Medical College, Valhalla, NY, USA. · Department of Otolaryngology-Head & Neck Surgery, Westchester Medical Center, Valhalla, NY, USA.

Observational · n=77

Key points

From the paper's conclusion
  • Intraoperative eSRT was feasible and yielded high response rates, particularly at apical and middle electrodes. However, intraoperative thresholds did not correlate with early postoperative values and cannot currently predict early postoperative eSRTs. These findings support the utility of intraoperative eSRT as a qualitative intraoperative tool, but further refinement is needed to improve its predictive value for postoperative cochlear implant programming.

Abstract

OBJECTIVE: To determine if intraoperative electrically evoked stapedial reflex thresholds (eSRT) correlate with postoperative eSRT and can predict programming levels in cochlear implant patients. DESIGN: Prospective cohort study conducted at a tertiary hospital. Seventy-seven patients who underwent cochlear implant surgery were included. Intraoperative and postoperative eSRT testing was performed at electrodes E1, E6, E12, E17, and E22, with postoperative testing conducted at a median of 4 weeks. RESULTS: Eighty-four cochlear implants in 77 patients met inclusion criteria, with intraoperative eSRT attempted in 72 ears and postoperative testing in 50. Intraoperative response rates were significantly higher than postoperative rates overall (83.1% vs. 70.9%, p = 0.0005), with the highest rates at middle and apical electrodes. Median intraoperative thresholds were significantly higher than postoperative thresholds (21.2 vs. 15.6, p < 0.0001), but no significant correlations were observed between the two time points. CONCLUSION: Intraoperative eSRT was feasible and yielded high response rates, particularly at apical and middle electrodes. However, intraoperative thresholds did not correlate with early postoperative values and cannot currently predict early postoperative eSRTs. These findings support the utility of intraoperative eSRT as a qualitative intraoperative tool, but further refinement is needed to improve its predictive value for postoperative cochlear implant programming.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers