Test-Retest Reliability of Myocene for Assessing Lower Limb Low-Frequency Fatigue: A Cross-Condition Reliability Study

Filipe Maia, Fábio Yuzo Nakamura, Sandro Tito, João Ribeiro

Universidade Politecnica · Universidade Federal de Pernambuco

Key points

From the paper's conclusion
  • The results demonstrated overall good (condition 1: 0.883 and 0.812; condition 2: 0.832 and 0.729 for dominant and non-dominant limb, respectively) and statistically significant (p < .001) reliability of LFF scores obtained using MyoceneⓇ, with SEM ranging from 1.1 (1,4%) to 1.9 (2.7%) and MDC values ranging from 3.03 (4.0%) to 5.27 (7.5%).

Abstract

This study aimed to assess the test-retest reliability of low-frequency fatigue (LFF) measured with MyoceneⓇ. Two conditions were tested: i) unmarked lower limbs assessed using the same set of electrodes (45 participants) and ii) marked anatomical points using a different set of electrodes (42 participants). The intra-class correlation coefficient (ICC) was calculated separately for each limb (dominant and non-dominant) under both conditions, along with the standard error of measurement (SEM), minimal detectable change (MDC), and Bland-Altman plots. The results demonstrated overall good (condition 1: 0.883 and 0.812; condition 2: 0.832 and 0.729 for dominant and non-dominant limb, respectively) and statistically significant (p < .001) reliability of LFF scores obtained using MyoceneⓇ, with SEM ranging from 1.1 (1,4%) to 1.9 (2.7%) and MDC values ranging from 3.03 (4.0%) to 5.27 (7.5%). In sum, MyoceneⓇ seems to reliably measure LFF, with slightly better results observed when using the same set of electrodes.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers