A short embedding protocol for diagnostic electron microscopy: long-term implementation and diagnostic outcomes in a high-volume clinical series.
Hui Zhou, Xin Jin Joseph Zhou
Electron Microscopy Laboratory, Quest Diagnostics Incorporated, Lewisville, TX, USA. · Renal Path Diagnostics/PathGroup & Department of Pathology, Baylor University Medical Center at Dallas, Dallas, TX, USA.
Key points
From the paper's conclusion- We evaluated whether reducing the polymerization step alone to 40 minutes at 120 °C, with upstream processing unchanged, could maintain diagnostic quality.
Abstract
Epoxy resin polymerization is a lengthy step in tissue preparation for transmission electron microscopy (TEM). We evaluated whether reducing the polymerization step alone to 40 minutes at 120 °C, with upstream processing unchanged, could maintain diagnostic quality. The protocol has been the sole routine method in a high-volume, accredited clinical TEM laboratory for more than 10 years. Among more than 30,000 consecutive cases, no embedding-related diagnostic inadequacy was identified in consultant pathologist sign-out records. In a prospective, blinded, split-specimen comparison of renal biopsies, both pathologists scored 4 or 5 on every feature for both methods. Nearly all scores were at the upper end of the scale, indicating comparable performance but not establishing statistical equivalence. Glomerular basement membrane thickness was closely similar (mean difference -12.7 nm, ~4%), within measurement variability and below diagnostic detection thresholds. Because curing behavior depends on the resin formulation and oven conditions, local validation is required before implementation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic utility of electron microscopy on frozen‑embedded tissues for renal biopsy pathology: a retrospective comparative study.International urology and nephrology · 2026
- Rapid femtosecond laser trenching accelerates cryo-electron tomography lamella workflowsbioRxiv · 2026
- Rapid post-mortem human temporal bone retrieval and cochlear preservation: a standardised protocol with systematic comparison of published methodologies.Annals of anatomy = Anatomischer Anzeiger : official organ of the Anatomische Gesellschaft · 2026
- Perfusion quality does not necessarily predict ultrastructural preservation after hyperosmotic brain perfusion.bioRxiv : the preprint server for biology · 2026
- Indium Tin Oxide Substrates Enable Coating-Free SEM Imaging and Simplified Preparation of Purified Fibrinogen Clots.Microscopy and microanalysis : the official journal of Microscopy Society of America, Microbeam Analysis Society, Microscopical Society of Canada · 2026
- Optimizing an ethylenediaminetetraacetic acid decalcification protocol for diagnostic bone biopsies: experience at an Academic Medical Center.Journal of histotechnology · 2026