Rapid femtosecond laser trenching accelerates cryo-electron tomography lamella workflows
C. M. Orr, P. N. M. Shah, E. C. Bragginton, S. Chakraborty, H. M. E. Duyvesteyn, S. Riggs, D. I. Stuart, A. Wagner
Key points
From the paper's conclusion- We present a pre-processing approach in which bulk material is removed by ultrashort-pulse laser ablation prior to milling, rapidly generating large trenches in vitrified grids under cryogenic conditions and leaving material to be finished by conventional milling to yield electron-transparent lamellae.
Abstract
Transmission electron microscopy requires electron-transparent specimens, typically below 200 nm thick. For cellular and tissue samples this is typically achieved by focused ion-beam (FIB) thinning to form lamellae. Milling times scale with specimen thickness, and high-pressure-frozen specimens more than 10 m thick can take up to 2 hours per lamella. We present a pre-processing approach in which bulk material is removed by ultrashort-pulse laser ablation prior to milling, rapidly generating large trenches in vitrified grids under cryogenic conditions and leaving material to be finished by conventional milling to yield electron-transparent lamellae. Laser ablation completes in seconds (4 seconds per pattern), compared with minutes to tens of minutes for FIB trench milling, and introduces negligible damage to the final lamellae. By replacing the most time-intensive, contamination-sensitive bulk-removal steps with seconds-long laser trenching, this open-source approach reduces time-to-lamellae, increasing throughput and preserves compatibility with standard cryo-FIB/SEM polishing and cryo-ET data-collection pipelines.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Femtosecond pulsed laser ablation for in-tissue electron cryo-tomographybioRxiv · 2026
- Adaptive Milling: imaging feedback-driven automated fabrication of consistently thin cryo-lamellaebioRxiv · 2026
- Optimized tissue sample preparation for cryo-electron tomography using serial lift-out.Biophysics reports · 2026
- Laser preparation for large‐scale FIB‐SEM tomography: Extended abstractce/papers · 2026
- Optical interference for the guidance of cryogenic focused ion beam milling beyond the axial diffraction limitNature Communications · 2026
- Making plant tissue accessible for cryo-electron tomographybioRxiv · 2026