Patient selection for leadless atrioventricular synchronous pacing: clinical and echocardiographic predictors of sustained device-derived AV synchrony with Micra AV.
Bart Jacobs, Thomas Van Weyenbergh, Alexandre Bohyn, Rik Willems, Alexander Van De Bruaene, Bert Vandenberk, Christophe Garweg
Department of Cardiovascular Sciences, University Hospital Leuven and KU Leuven, Leuven, Belgium. · Department of Public Health and Primary Care, KU Leuven, Leuven Biostatistics and Statistical Bioinformatics Centre (L-BioStat), Leuven, Belgium.
Key points
From the paper's conclusion- Clinical and echocardiographic characteristics may help identify patients most likely to achieve sustained atrioventricular coupling with Micra AV. Markers of preserved atrial function, limited atrial remodelling, and lower filling pressures may further refine patient selection.
Abstract
BACKGROUND: The Micra AV leadless pacemaker allows atrioventricular synchronous pacing through accelerometer-based sensing of atrial mechanical activity (A4 amplitude). However, determinants of sustained atrioventricular synchrony (AVS) in real-world conditions remain poorly defined. OBJECTIVE: To identify clinical and echocardiographic predictors of device-derived AVS 1 year after Micra AV implantation, and to assess the association between 1-year AVS and baseline A4 amplitude, and predictors of A4 amplitude. METHODS: In patients undergoing Micra AV implantation, baseline clinical and echocardiographic data were collected. One-year AVS was evaluated in patients with ≥80% ventricular pacing using a device-derived index reflecting atrial mechanical sensing. Predictors were identified using MI-LASSO and stepwise regression with bootstrap validation. RESULTS: 94 patients were included; high-degree AV block was the primary pacing indication (N = 88). At 1 year, 38 patients had ≥80% ventricular pacing; median AM-VP% was 82.9% (IQR 73.8-87.8) and AVS index was 83.1% (IQR 77.0-87.7). Predictors of device-derived 1-year AVS included older age, higher left atrial reservoir strain rate, lower right atrial volume index, and lower E/e'. For AVS, R2 was 0.80 (MI-LASSO) and 0.55 (stepwise). One-year AVS was strongly associated with baseline A4 amplitude (p < 0.001). Predictors of A4 amplitude included hypertension, absence of prior cardiac surgery, higher TAPSE, higher left atrial reservoir strain and preserved left atrial conduit strain rate. CONCLUSIONS: Clinical and echocardiographic characteristics may help identify patients most likely to achieve sustained atrioventricular coupling with Micra AV. Markers of preserved atrial function, limited atrial remodelling, and lower filling pressures may further refine patient selection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Long-Term Performance of Single-Chamber Atrioventricular-Synchronous Leadless Pacing: Results From 4 Years of Serial Follow-up.Journal of cardiovascular electrophysiology · 2026
- MicraAV Versus Dual-Chamber Transvenous Pacemakers in Patients with Atrioventricular Block Following Transcatheter Aortic Valve Replacement: A 12-Month Single-Center Retrospective Study.Journal of clinical medicine · 2026
- Predictors of Heart Failure Hospitalization in Patients with the First Marketed Leadless Pacemaker -Micra VR.Indian pacing and electrophysiology journal · 2026
- First-in-Human Permanent Totally Leadless Left Ventricular Septal Endocardial Pacing With Atrioventricular Synchrony.JACC. Case reports · 2026
- Long-term outcomes of single atrial lead pacing in sick sinus syndrome: risk stratification using wenckebach block rateEuropean Heart Journal · 2026
- Multicenter Study of the Electrical Characteristics and Short-Term Outcomes of the Aveir VR Leadless Pacemaker.Journal of the American Heart Association · 2026