The Review of scientific instruments

Design and experimental evaluation of a small-size, high-precision nuclear magnetic resonance probe for enlarged magnetic field measurement region on small-bore magnetic resonance imaging magnets.

Guyue Zhou, Wenchen Wang, Yiqing Yin, Haotian Miao, Zhan Liu, Shufeng Wei, Wenhui Yang, Yunxing Song,

Wuhan National High Magnetic Field Center, Huazhong University of Science and Technology, Wuhan 430074, People's Republic of China. · School of Information Technology and Electrical Engineering, The University of Queensland, St. Lucia, Brisbane, QLD 4072, Australia.

Key points

From the paper's conclusion
  • These results demonstrate that the proposed microprobe provides a low-cost, stable, high-precision, and broadly applicable solution for extending accessible magnetic field mapping in bore-constrained MRI magnets.

Abstract

High-quality magnetic resonance imaging (MRI) relies on accurate magnetic field characterization; however, commercial nuclear magnetic resonance (NMR) magnetic field probes are often limited in boundary-accessible field mapping for bore-constrained magnets. In this study, a miniature free-induction-decay-based NMR magnetic field probe was developed, integrating a solenoid microcoil, a compact tuning/matching network, and a liquid sample unit, with a maximum width of ∼8 mm. Compared with a commercial NMR reference probe, the developed microprobe showed close agreement in single-point validation and yielded a mean absolute deviation of 0.27 μT in axial comparison measurements (corresponding to 0.039 ppm at 7 T), confirming its field-measurement accuracy and spatial tracking capability. Temperature-sensitivity tests over a range of 15-35 °C showed that the repeated-measurement standard deviation remained within 0.019-0.046 ppm, indicating good measurement stability. In wall-adjacent measurements inside a 172 mm-bore magnet, the maximum accessible trajectory diameter increased from 152 to 164 mm, corresponding to a 7.9% improvement in an accessible mapping range. Spherical-surface mapping of a 50 mm diameter spherical volume further yields a mean absolute deviation of 0.143 μT and a Pearson correlation coefficient greater than 0.99. These results demonstrate that the proposed microprobe provides a low-cost, stable, high-precision, and broadly applicable solution for extending accessible magnetic field mapping in bore-constrained MRI magnets.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Design and experimental evaluation of a small-size, high-precision nuclear magnetic resonance probe for enlarged magnetic field measurement region on small-bore magnetic resonance imaging magnets — Science Skim