Design and experimental evaluation of a small-size, high-precision nuclear magnetic resonance probe for enlarged magnetic field measurement region on small-bore magnetic resonance imaging magnets.
Guyue Zhou, Wenchen Wang, Yiqing Yin, Haotian Miao, Zhan Liu, Shufeng Wei, Wenhui Yang, Yunxing Song,
Wuhan National High Magnetic Field Center, Huazhong University of Science and Technology, Wuhan 430074, People's Republic of China. · School of Information Technology and Electrical Engineering, The University of Queensland, St. Lucia, Brisbane, QLD 4072, Australia.
Key points
From the paper's conclusion- These results demonstrate that the proposed microprobe provides a low-cost, stable, high-precision, and broadly applicable solution for extending accessible magnetic field mapping in bore-constrained MRI magnets.
Abstract
High-quality magnetic resonance imaging (MRI) relies on accurate magnetic field characterization; however, commercial nuclear magnetic resonance (NMR) magnetic field probes are often limited in boundary-accessible field mapping for bore-constrained magnets. In this study, a miniature free-induction-decay-based NMR magnetic field probe was developed, integrating a solenoid microcoil, a compact tuning/matching network, and a liquid sample unit, with a maximum width of ∼8 mm. Compared with a commercial NMR reference probe, the developed microprobe showed close agreement in single-point validation and yielded a mean absolute deviation of 0.27 μT in axial comparison measurements (corresponding to 0.039 ppm at 7 T), confirming its field-measurement accuracy and spatial tracking capability. Temperature-sensitivity tests over a range of 15-35 °C showed that the repeated-measurement standard deviation remained within 0.019-0.046 ppm, indicating good measurement stability. In wall-adjacent measurements inside a 172 mm-bore magnet, the maximum accessible trajectory diameter increased from 152 to 164 mm, corresponding to a 7.9% improvement in an accessible mapping range. Spherical-surface mapping of a 50 mm diameter spherical volume further yields a mean absolute deviation of 0.143 μT and a Pearson correlation coefficient greater than 0.99. These results demonstrate that the proposed microprobe provides a low-cost, stable, high-precision, and broadly applicable solution for extending accessible magnetic field mapping in bore-constrained MRI magnets.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An integrated NMR field probe using rapid non-adiabatic field switch-off for (ultra) low field MRI.Communications engineering · 2026
- Compact magnet design for magnetic resonance: A system-aware overview of optimization strategies, manufacturing technologies, and system-level trade-offs.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2026
- The world's highest field optical microscopy probe compatible with 42 T steady magnetic field and down to 1.8 K.The Review of scientific instruments · 2026
- Narrowband isotropic magnetic particle imaging with FFL-parallel excitation.Physics in medicine and biology · 2026
- From Repeatability to Reproducibility in Quantitative Low-Field MRI: A Metrology-Informed Framework for Phantom Evidence.Journal of imaging · 2026
- Magnetic Imaging with High Spatial Resolution Using a Single Niobium-Nitride-Based Three-Dimensional Nanobridge JunctionarXiv · 2026