The Review of scientific instruments

Cyclotron-frequency stability via single-ion fluorescence in a Penning trap with a cryogen-free superconducting magnet.

A Schupeta, D Yousaf, M Almagro, M Hurtado, J M Palomino, J Berrocal, M Block, Ch E Düllmann,

Departamento de Física Atómica, Molecular y Nuclear, Universidad de Granada, 18071 Granada, Spain. · Ultrasonics Laboratory, Universidad de Granada, 18071 Granada, Spain.

Key points

From the paper's conclusion
  • The short-term stability was investigated using the FD-FT-ICR technique, resulting in a minimum relative magnetic-field variation of δB/B ≃ 5 × 10-8 for averaging times between 30 and 60 s.

Abstract

In this work, we investigate the magnetic-field stability of a Penning trap operated with a cryogen-free superconducting magnet through cyclotron-frequency measurements of a single laser-cooled calcium ion using two complementary photon-based detection techniques: a pulsed optical method for determining the three ion eigenfrequencies and Fluorescence-Detected Fourier-Transform Ion-Cyclotron-Resonance (FD-FT-ICR). For the former technique, the data-analysis procedure was revisited, considering it as the dominant contribution to the measurement uncertainty and yielding a long-term magnetic-field drift (1/B)(dB/dt) = -3.07(32) × 10-9 h-1. This value is comparable with those reported for high-precision Penning-trap experiments employing liquid-helium-based superconducting magnets. The short-term stability was investigated using the FD-FT-ICR technique, resulting in a minimum relative magnetic-field variation of δB/B ≃ 5 × 10-8 for averaging times between 30 and 60 s. Furthermore, this technique enables direct cyclotron-frequency determinations on timescales as short as a few seconds, providing access to magnetic-field fluctuations that are generally not resolved in conventional Penning-trap experiments.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Cyclotron-frequency stability via single-ion fluorescence in a Penning trap with a cryogen-free superconducting magnet — Science Skim