Low Vibration Closed-Cycle Flow Cryostat for Ion-Cavity Experiments
Andrew Laugharn
Key points
From the paper's conclusion- We have developed a custom 4 K cryostat cooled by a ColdEdge Technologies (CET) Stinger closed-cycle $^{4}\text{He}$ flow cryogenic system to study various ion trap optical cavity structures designed to serve as a photonic interconnect between remote, compute-capable trapped $^{171}\text{Yb}^{+}$ ion quantum processors, enabling long baseline entanglement distribution $(>10\text{ km})$.
Abstract
Cryogenic operation below 10 K suppresses trap-electrode and dielectric heating of trapped ions while cryopumping residual gases to ultra-high vacuum (UHV) without requiring a bakeout. These advantages are particularly important for cavity-integrated ion traps, where nearby dielectric surfaces can increase motional heating and rapid design-build-test iterations benefit from eliminating bakeout cycles. We have developed a custom 4 K cryostat cooled by a ColdEdge Technologies (CET) Stinger closed-cycle $^{4}\text{He}$ flow cryogenic system to study various ion trap optical cavity structures designed to serve as a photonic interconnect between remote, compute-capable trapped $^{171}\text{Yb}^{+}$ ion quantum processors, enabling long baseline entanglement distribution $(>10\text{ km})$. We integrate an Alpine Quantum Technologies (AQT) Paul trap and Fabry-P\'erot optical cavity at the 4 K stage. The cooling capacities of the cryostat 40 K and 4 K stages were measured to be 5 W at 45 K and 0.5 W at 6.1 K. Vibrations of the 4 K stage relative to the optical table are less than $12\text{ nm}_{\text{RMS}}$ in three dimensions, as measured using a Michelson interferometer. An ultra-high vacuum (UHV) is inferred by studying the rate of hydrogen-$\mathrm{Yb}^{+}$ ion collisions. Ion storage times exceeding 5 days without laser cooling have been reported and the position of the trapped ions are reproducible within $\sim10\text{ }\mu\text{m}$ following repeated thermal cycles to room temperature. The uptime of this system is 72 days and offers convenient access to the 4 K stage for in-vacuum maintenance and system modifications. Using this platform, we demonstrate spatial overlap and interaction between a trapped $^{174}\text{Yb}^{+}$ ion and the optical cavity mode on the 1345 nm transition, establishing preliminary experimental validation of the integrated cryogenic ion-cavity structure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A room-temperature extreme high vacuum system for trapped-ion quantum information processingQuantum Science and Technology · 2026
- Ion trap on borosilicate substrate with integrated femtosecond-laser-written waveguideQuantum Science and Technology · 2026
- Cancellation of cryocooler-induced vibration in trapped-ion systemsarXiv · 2026
- A low-volume cold-atom microwave clock with an additively manufactured cavityarXiv · 2026
- Integrated-photonics-based systems for polarization-gradient cooling of trapped ionsLight Science & Applications · 2026
- Multi-hour stable trapping and threshold steady state of non-laser-coolable ions with ultracold atoms in a hybrid traparXiv · 2026