The Review of scientific instruments

Evaluation of the high-performance TechnoTrak CR-39 for detection of 1-9 MeV fusion protons.

N Vanderloo, A T Lanzrath, E Edwards, T E Evans, Z Tessema, M Cufari, S G Dannhoff, B C Foo,

Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, USA.

Key points

From the paper's conclusion
  • The results indicate that TechnoTrak CR-39 detectors exhibit high detection efficiency over a wide proton energy range (∼1-5 MeV), remain sensitive to protons up to ∼9 MeV, and have lower intrinsic noise than widely used TASTRAKTM CR-39 produced by Track Analysis Systems Ltd.

Abstract

In this paper, the MeV-proton response of a high-sensitivity, low-noise CR-39 nuclear track detector, TechnoTrak, produced by the Chiyoda Technol Corporation, is evaluated. Plastic CR-39 nuclear track detectors are commonly used for precise time-integrated measurement of charged particles. Such measurements are crucial for diagnosing inertial confinement fusion (ICF) and high energy density (HED) experiments at the Omega Laser Facility, the Z Pulsed Power Facility, and the National Ignition Facility. The response of TechnoTrak CR-39 detectors was tested with protons produced from DD and D3He fusion reactions at the MIT linear ion accelerator. The results indicate that TechnoTrak CR-39 detectors exhibit high detection efficiency over a wide proton energy range (∼1-5 MeV), remain sensitive to protons up to ∼9 MeV, and have lower intrinsic noise than widely used TASTRAKTM CR-39 produced by Track Analysis Systems Ltd. This wide detectable energy range and low noise floor may benefit future HED and ICF experiments requiring low background levels and broad proton energy sensitivity.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Evaluation of the high-performance TechnoTrak CR-39 for detection of 1-9 MeV fusion protons — Science Skim