Upgrades to the National Ignition Facility neutron time-of-flight diagnostic for operation at yields greater than 10 MJ.

S Kerr, K D Hahn, J Jeet, M Eckart, A S Moore, E Mariscal, R Hatarik, H Khater,

Lawrence Livermore National Laboratory, Livermore, California 94550, USA.

Key points

From the paper's conclusion
  • We present two upgrades to the NIF neutron time-of-flight (nToF) diagnostic suite that will enable measurements at high yields.

Abstract

The National Ignition Facility (NIF) is reaching MJ fusion yields and has plans to reach tens of MJ. Diagnostic operation at these yields introduces challenges, from detector linearity at high signal levels to increased backgrounds. We present two upgrades to the NIF neutron time-of-flight (nToF) diagnostic suite that will enable measurements at high yields. First, silver neutron attenuators are being installed in the collimated nToF beamlines, reducing neutron flux and maintaining optimal signal-to-background levels. Silver was selected for its smooth neutron transmission cross section and sufficient γ-ray transmission to retain the gamma peak used for hot-spot velocity measurements. Each line-of-sight will have four insertable silver pucks, providing ∼2-20× attenuation of 14 MeV neutrons. Second, microchannel plate (MCP) photomultiplier tubes using Zener diodes to establish constant MCP-to-anode acceleration voltages have been installed in the quartz Cherenkov detectors. These provide stable instrument response functions across large gain ranges and permit operation at MCP gains near unity. The combined upgrades will allow NIF nToFs to function at all projected future yields.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Upgrades to the National Ignition Facility neutron time-of-flight diagnostic for operation at yields greater than 10 MJ — Science Skim