The Review of scientific instruments

Nanosecond-resolved 266 nm Mach-Zehnder interferometry of laser-produced dense plasmas generated in supercritical helium.

Kyusang Cho, Juho Lee, Gunsu Yun

Department of Physics, POSTECH, 77 Cheongam-ro, Nam-gu, Pohang 37673, Republic of Korea.

Key points

From the paper's conclusion
  • We developed a nanosecond-resolved 266 nm Mach-Zehnder interferometer for dense laser-produced plasmas generated in supercritical helium at 100 bar.

Abstract

We developed a nanosecond-resolved 266 nm Mach-Zehnder interferometer for dense laser-produced plasmas generated in supercritical helium at 100 bar. A 1064 nm pump was focused into the helium, and the induced phase shift of a 266 nm probe was recorded with a 3 ns intensified charge-coupled device gate. Separately acquired plasma-arm-only and reference-arm-only images were used to normalize each interferogram before two-dimensional Fourier phase retrieval. Owing to the inherent limitations of single-wavelength interferometry, which cannot distinguish between free-electron and neutral-species contributions, the measured phase shift is reported as an electron-density-equivalent quantity using an electron-only conversion. At 10 ns, the maximum line-of-sight-integrated value in the averaged frame was 2.01 × 1017 cm-2, whereas the median from 19 fringe-recoverable frames was 2.22 × 1017 cm-2. A regularized finite-shell Abel inversion was applied to the 10 ns averaged profile. Varying the emission-defined outer-radius criterion from 10% to 50% produced on-axis electron-density-equivalent values of 2.04-2.42 × 1018 cm-3, quantifying the Abel-inverted on-axis sensitivity to the boundary definition. Adopting the 30% criterion as the nominal boundary gave (2.31 ± 0.16) × 1018 cm-3 from phase-noise uncertainty, excluding unresolved systematic effects. Normalized visibility, Fourier-sideband signal-to-noise ratio, phase noise, and region of interest quantify the region over which phase retrieval remains well conditioned. Electron-ion free-free absorption and finite-collision-frequency effects are estimated, whereas probe refraction remains an unresolved systematic uncertainty. Our UV interferometer demonstrates the integrated implementation and quantitative characterization of the diagnostic in 100 bar supercritical helium, while defining the limits of its absolute-density interpretation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Nanosecond-resolved 266 nm Mach-Zehnder interferometry of laser-produced dense plasmas generated in supercritical helium — Science Skim