Development and commissioning of a LaBr3:Ce, Sr-based collimated hard x-ray spectrometer for MAST-U.
Z Mianowska, P Beaumont, M Beldishevski, R de Chazal, C Downing, Z Ghani, B Kandan, V G Kiptily,
(UKAEA) United Kingdom Atomic Energy Authority, Culham Campus, Abingdon, Oxfordshire, OX14 3DB, United Kingdom. · University of Birmingham, School of Physics and Astronomy, Birmingham, West Midlands, B15 2TT, United Kingdom.
Key points
From the paper's conclusion- Following installation, the collimated HXR diagnostic demonstrated reliable operation by recording gamma-ray events and hard x-ray bursts during both ohmic and neutral beam injection heated plasma discharges, confirming the successful commissioning of the diagnostic on MAST-U.
Abstract
A new collimated hard x-ray (HXR) spectrometer has been developed and installed on the Mega Amp Spherical Tokamak Upgrade (MAST-U) to support future studies of runaway electron generation and provide a diagnostic capability to plasma-facing components monitoring. The diagnostic is based on an Sr-codoped LaBr3:Ce scintillator (2″ × 2″) coupled to a fast data acquisition system capable of recording raw scintillation waveforms in both pulse and continuous modes, allowing detailed reconstruction of individual events using algorithms for optimization of spectrometric parameters and pulse-shape discrimination. The detector was energy-calibrated up to 4.8 MeV using standard spectroscopy sources and a 66Ga radioisotope. Its performance was then evaluated at the MC40 cyclotron (University of Birmingham), demonstrating stable operation at 1.1 Mcps, with gain shifts of less than 4%. Following successful validation, the diagnostic was installed using an existing midplane line-of-sight configuration of the MAST-U, passing through both plasma-facing components and the central column region. A neutron flux map was simulated to estimate the radiation environment within the diagnostic field of view defined by a 15 cm diameter penetration through 1.5 m of concrete biological shielding acting as a collimator, together with an additional 10 cm polyethylene layer. Following installation, the collimated HXR diagnostic demonstrated reliable operation by recording gamma-ray events and hard x-ray bursts during both ohmic and neutral beam injection heated plasma discharges, confirming the successful commissioning of the diagnostic on MAST-U.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance Assessment of a GEM-Based SXR Diagnostic System Operating in the WEST Plasma EnvironmentEnergies · 2026
- Absolutely calibrated scintillator stack spectrometer for high-intensity laser-plasma experiments.The Review of scientific instruments · 2026
- Photodiode based multi-modal diagnostic for low-energy neutral beam injection in the LTX-β spherical tokamak.The Review of scientific instruments · 2026
- Development of a fusion proton diagnostic for the Wisconsin High Temperature Superconductor (HTS) Axisymmetric Mirror (WHAM).The Review of scientific instruments · 2026
- High-energy gamma-ray spectrometry with LYSO:Ce scintillators for fusion plasma experiments.The Review of scientific instruments · 2026
- Broad-energy characterization of MAPD-3NM2 photodiode Matrix coupled with LaBr3(Ce)and YAG:Ce scintillators for space ApplicationsBAKU STATE UNIVERSITY JOURNAL of PHYSICS & SPACE SCIENCES · 2026