Conceptual design of the electron-optics of the magnetic electron recoil gamma-ray spectrometer.
J H Kunimune, S Huber, S Mackie, E Panontin, N Fensterheim, M Dessena, E Gallo, F Montanari,
Plasma Science and Fusion Center, Massachusetts Institute of Technology, 167 Albany St., Cambridge, Massachusetts 02139, USA. · Eni S.p.A., Piazzale Mattei 1, Rome, Italy.
Key points
From the paper's conclusion- We propose that a magnetic electron recoil gamma-ray spectrometer (MERGS) can make gamma-ray measurements in these high-background environments.
Abstract
In nuclear fusion reactors, deuterium and tritium are fused together. Most reactions produce a 4He ion and a 14 MeV neutron, while about 2 × 10-5 of them produce a 5He ion and a gamma photon. One of the most important measurements from a fusion reactor is the total fusion power, which can be measured using the 14 MeV neutrons or the gamma-rays. Using both provides two independent measurements, which is valuable for researchers and regulators. Facilities such as JET use scintillators such as LaBr3 to measure gamma-ray emission, but such detectors could struggle with background on high-power reactors such as SPARC and ITER. We propose that a magnetic electron recoil gamma-ray spectrometer (MERGS) can make gamma-ray measurements in these high-background environments. MERGS will consist of a foil that converts photons to electrons via Compton scattering, a magnetic electron-optic system that focuses and disperses the electron beam, and a detector array that counts the electrons and uses their positions to infer their energy. A conceptual design framework of MERGS is presented, along with three numerically optimized electron-optic geometries for three different resolution requirements. Simulations are presented showing that each presented design can meet its requirements. This shows that MERGS can measure the gamma-ray emission spectrum more effectively than scintillators for a variety of price points, enabling improved measurements of fusion power as well as the study of fast ion physics via other nuclear reactions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Gamma-electron conversion on MERGS (magnetic electron recoil gamma-ray spectrometers) for fusion diagnostics.The Review of scientific instruments · 2026
- The next-generation magnetic recoil spectrometer for measurements of the absolute neutron spectrum at Pacific Fusion's demonstration system.The Review of scientific instruments · 2026
- Operational domain estimation of a gamma-ray spectrometer for deuterium–tritium fusion power measurement at ITERPlasma Physics and Controlled Fusion · 2026
- Magnetic recoil neutron spectrometry for inertial confinement fusion: To ignition and beyond (invited).The Review of scientific instruments · 2026
- A compact recoil spectrometer for time integrated neutron energy spectrum measurements on the Z machine (invited).The Review of scientific instruments · 2026
- Impact of instrumental effects and target thickness on the SPARC magnetic proton recoil spectrometer response function.The Review of scientific instruments · 2026