Expanded higher energy channel for the National Ignition Facility's soft x-ray opacity spectrometer (OpSpec).
M S Wallace, R F Heeter, Y P Opachich, A E Peterson, E C Dutra, P Cho, D C Mayes, B A Hobbs,
Nevada National Security Site, Livermore Operations, Livermore, California 94550, USA. · Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
Key points
From the paper's conclusion- The National Ignition Facility's soft x-ray opacity spectrometer (OpSpec) currently records time-integrated spectra from ∼550 to 2000 eV.
Abstract
The National Ignition Facility's soft x-ray opacity spectrometer (OpSpec) currently records time-integrated spectra from ∼550 to 2000 eV. To accomplish this, two elliptical acid-phthalate crystals with different shapes are used together. They are called the low energy and standard energy OpSpec channels and they span roughly 550-1000 and 1000-2000 eV, respectively. Replacing the crystal in the low-energy mount with a PET crystal gives a new higher-energy channel. Slight modifications to the shape of the ellipse give the new channel an energy range of 1750-3000 eV. The new geometry, testing results, and data from the NIF opacity campaign are discussed in this article.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- First results from the time resolved opacity spectrometer on the National Ignition Facility (invited).The Review of scientific instruments · 2026
- Characterization of a high-efficiency, streaked x-ray spectrometer for warm-dense matter studies on the National Ignition Facility (invited).The Review of scientific instruments · 2026
- A proposal for a 1D-imaging broadband x-ray crystal spectrometer enabled by a sagittally focusing x-ray mirror.The Review of scientific instruments · 2026
- The OMEGA-extended performance streaked optical pyrometer with spectral resolution.The Review of scientific instruments · 2026
- AlMg transmission measurements on the National Ignition Facility with novel uncertainty quantification method for point-projection geometryPhysics of Plasmas · 2026
- Design and characterization of a hohlraum for dynamic radiography at the OMEGA EP (extended performance) laser system.The Review of scientific instruments · 2026