Digital least-squares amplitude detection for high-speed atomic force microscopy.
Suresh Babu Lavanya, Yi-Chih Lin
Department of Chemistry, University of Texas at Austin, Austin, Texas 78712, USA.
Key points
From the paper's conclusion- These results establish digital LSQ as an additional amplitude-detection strategy for HS-AFM, particularly for applications involving noisy cantilever deflection signals where tunable control of the response-time/amplitude-fluctuation trade-off is needed.
Abstract
High-speed atomic force microscopy (HS-AFM) in amplitude-modulated (tapping) mode requires rapid and reliable amplitude detection for real-time feedback control. Here, we develop a low-latency field-programmable gate array (FPGA)-based digital least-squares (LSQ) amplitude detector for HS-AFM operation. Although LSQ is a well-established signal-processing method, the novelty of this work lies in its implementation as a high-frequency, microsecond-response amplitude detector for real-time HS-AFM feedback. To enable real-time operation, we implemented a two-parameter LSQ formulation on FPGA hardware, in which the adjustable fitting-window length defines the number of sampled points used for each amplitude estimate. We evaluated LSQ using numerical simulations, signal-generator measurements, HS-AFM deflection-signal analysis, and biological HS-AFM imaging. Under matched response-time conditions, numerical simulations showed that LSQ produced amplitude-estimate fluctuations comparable to Fourier-amplitude-based detection and lower than those produced by sample-and-hold-based detection (SHB) and phase-shift-based detection. Experimental measurements further showed that shorter LSQ windows provided SHB-like response times, whereas longer windows reduced amplitude-estimate fluctuations. LSQ also supported stable HS-AFM imaging of annexin-V lattices on supported lipid bilayers. These results establish digital LSQ as an additional amplitude-detection strategy for HS-AFM, particularly for applications involving noisy cantilever deflection signals where tunable control of the response-time/amplitude-fluctuation trade-off is needed.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optical-feedback scanning ion conductance microscopy with high-speed and low-overshoot probe control.Micron (Oxford, England : 1993) · 2026
- Dislocation and distortion correction using back-EMF-derived line-data information in compact resonant electromagnetic MEMS scanning confocal microscopy.Applied optics · 2026
- Pushing the limits of atomic force microscopy: Emerging methods for speed, resolution, and quantitative analysis.Journal of microscopy · 2026
- Low-latency axial depth estimation of 2.82 ms using light field microscopy.Optics express · 2026
- A high signal-to-noise ratio and high-frequency seesaw cantilever for high-speed atomic force microscopyNature Communications · 2025
- Cross-scale high-bandwidth atomic force microscopy with a stick-slip nanopositionerNature Communications · 2025