The Review of scientific instruments

Development of a multichannel filterscope system for two-dimensional plasma imaging in a magnetic mirror.

Dongha Kim, M Lee, D Oh, Y Jung, C Moon, S G Lee, G B Ko, Junghee Kim,

Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea. · Korea Institute of Fusion Energy, Daejeon 34133, Republic of Korea.

Key points

From the paper's conclusion
  • We have developed a multichannel filterscope system for time-resolved two-dimensional (2D) plasma emission imaging in the KAIST magnetic mirror device.

Abstract

We have developed a multichannel filterscope system for time-resolved two-dimensional (2D) plasma emission imaging in the KAIST magnetic mirror device. The system consists of three compact in-vessel receiver modules with slit-based pinhole-camera geometry, an optical feedthrough coupled to fiber bundles, and external silicon photomultiplier-based detection units. The receiver modules, separated by 120°, provide 21 line-integrated viewing chords across the plasma cross section. Plasma emission is spatially selected by a precision slit, collimated by lens optics, spectrally filtered by an interchangeable narrow bandpass filter, and transmitted outside the vacuum vessel through the optical feedthrough and fiber bundles. Initial measurements during helium discharges successfully exhibited simultaneous time-resolved detection of neutral helium emission near 668 nm from all 21 channels. The gain-corrected line-integrated signals were then used to reconstruct 2D emissivity distributions. The reconstructed images showed a non-axisymmetric emission structure and its temporal evolution, demonstrating the capability of the developed system for time-resolved 2D plasma emission imaging.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers