The Review of scientific instruments

A thermopile array and stereo vision fusion system for small-tube temperature sensing.

Dexin Hou, Tiancheng Xie, Yiming Lv, Ji Mi, Jiong Ding, Dongfang Hu

College of Metrology Measurement and Instrument, China Jiliang University, Hangzhou, China.

Key points

From the paper's conclusion
  • These results demonstrate that multi-sensor fusion effectively overcomes the size-of-source effect in low-cost infrared sensing.

Abstract

This paper presents a non-contact temperature measurement scheme for small-diameter tubes by fusing a low-resolution thermopile array (TPA) with visible-light stereo vision. Stereo vision locates the tube in three dimensions. Then, this geometry is projected onto the TPA imaging plane to constrain the radiation integration model. A spherical calibration target with an internal heat source is proposed, achieving a sub-pixel accuracy of 1/28 of a pixel via symmetry-based active aiming. Experiments on 3 and 2 mm heated tubes show that the proposed method reduces the mean absolute percentage error from 28% and 36% to 3.4% and 4.3%, respectively. Furthermore, the system resolves emissivity-induced circumferential temperature non-uniformities with a rotational resolution of 5°. These results demonstrate that multi-sensor fusion effectively overcomes the size-of-source effect in low-cost infrared sensing.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A thermopile array and stereo vision fusion system for small-tube temperature sensing — Science Skim