Interferometric diameter topography calibration of test spheres
Tillman Neupert-Wentz, Guido Bartl, René Schödel
Key points
From the paper's conclusion- The uncertainty budget for a diameter is presented, with a combined standard uncertainty of 20 nm.
Abstract
Test spheres as used in the calibration of coordinate measuring machine stylus and other free form measuring devices are typically characterised by a small set of diameters and the sphericity along these diameters. This sparse sampling limits the number of positions where the true diameter is known, reducing the sphere’s utility in calibration. To enhance data density and decrease the uncertainty of calibrations, an extension to PTB’s double-ended interferometer (DEI) (Rau and Schödel 2014 Double-Ended Interferometer for Measuring Gauge Blocks Without Wringing pp 859–62) has been developed (Schödel and Fischedick 2021 Meas. Sci. Technol. 32 084010). This work presents the implementation, measurements and evaluation methods for this extension. First, the DEI including the extension is introduced and the optical setup explained. Then the evaluation methods are presented. The evaluation is performed in two steps. First, pixelwise diameter values are calculated for all measurements performed on a sphere, then, a stitching algorithm is employed to combine a series of measurements to one representation of the sphere surface figure. The evaluation is tested with a measurement series on a 30 mm silicon sphere. The uncertainty budget for a diameter is presented, with a combined standard uncertainty of 20 nm.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- First high-precision characterisation of the absolute diameter and complete topography of a 30 mm silicon sphereMeasurement Science and Technology · 2026
- Uncertainty analysis of sphere parameter estimation from NMM-1 point cloud measurementstm - Technisches Messen · 2026
- Absolute surface metrology using gradient descent-optimized multi-surface phase-shifting interferometry for interferometer systematic error calibration.Optics express · 2026
- Subaperture stitching measurement for optical components based on the trust region reflective algorithm.Optics express · 2026
- High-accuracy synchronous distortion-compensated stereo phase measuring deflectometry with orthogonal geometric constraintsOptics Express · 2026
- Ronchi-calibrated bidirectional fast volume-repeat OCT for high-speed 4D imaging.Optics letters · 2026