A system for automated measurements of anisotropic elastic moduli using resonant ultrasound spectroscopy.
Joshua Ward, Matthew Cherry, Micheal Burba, Mark Obstalecki
University of Dayton Research Institute, Dayton, Ohio 45469, USA. · Air Force Research Laboratory, Materials and Manufacturing Directorate, Wright-Patterson AFB, Dayton, Ohio 45469, USA.
Key points
From the paper's conclusion- We present a novel automated system designed to streamline the RUS measurement process.
Abstract
Resonant Ultrasound Spectroscopy (RUS) is a powerful, non-destructive technique for precisely determining a material's elastic properties by measuring resonant frequencies and extracting elastic constants via an inversion algorithm. This information is crucial for understanding material behavior under various conditions and is applicable in diverse fields ranging from materials science and engineering to geophysics and aerospace. Traditional RUS is often time-consuming and requires significant expertise. We present a novel automated system designed to streamline the RUS measurement process. The system incorporates automated specimen positioning, excitation, signal acquisition, and data analysis, enhancing measurement efficiency and reproducibility. A key feature enabling robust automation is the integration of mode shape information into the inversion algorithm. Traditional frequency-matching RUS can be susceptible to convergence toward a local minimum, particularly for complex or highly damped materials. By incorporating mode shapes as constraints, our approach improves algorithm convergence and robustness. System performance and baseline accuracy are first demonstrated through benchmark measurements on the common titanium alloy Ti-6Al-4V, with RUS results matching pulse-echo ultrasound and mechanical tensile testing. Furthermore, to demonstrate the system's capability on elastically anisotropic materials without an a priori initial guess, measurements were performed on a high-purity single-crystal zinc specimen with its crystallographic c-axis aligned with a principal dimension. The automated inversion successfully recovered all five independent hexagonal elastic constants (C11, C33, C44, C12, and C13) in close agreement with literature values, confirming the robustness of the mode-shape-correlated approach for anisotropic media.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Determination of anisotropic elastic constants of additively manufactured Ti–6Al–4V with complex geometryJournal of Materials Research and Technology · 2026
- Fixed-phase resonance tracking for fast Nonlinear Resonant Ultrasound Spectroscopy.Ultrasonics · 2026
- Optimization of Ultrasonic Parameters and Model Development for Nondestructive CTE Measurement of LAS Ultra-Low-Expansion Glass Ceramics.Materials (Basel, Switzerland) · 2026
- Real-Time Laser Focus Tracking for Compressive Testing: Modular Rust Architecture with OPC UA and Web GUI for Laser Ultrasonics.Sensors (Basel, Switzerland) · 2026
- i-Rheo-Indent: Broadband rheological characterisation of soft solids by macroscale indentation.Rheologica acta · 2026
- Amplitude attenuation study and prediction model for large-load ultrasonic vibration systems: revealing the multi-factor coupling attenuation mechanism and stabilization strategies.Ultrasonics · 2026