Preliminary Results of the Use of a Robotic System in Spine Surgery.
Helton Luiz A Defino, Vinicius Marques Carneiro, Matheus Pippa Defino, Rian Souza Vieira, Thiago Dorigão, Denylson Sanches Fernandes
Department of Orthopedics and Anesthesiology, Faculdade de Medicina de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, SPBrazil. · Neurosurgery Division, Department of Surgery and Anatomy, Faculdade de Medicina de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, SPBrazil.
Key points
From the paper's conclusion- The robotic system demonstrated high accuracy in implant placement and, with increasing experience in its use, was associated with reductions in fluoroscopy time, intraoperative radiation exposure, and the time to record the patient's anatomy.
Abstract
OBJECTIVE: To evaluate the preliminary results of the use of the robotic system in spinal surgery. METHODS: The current prospective study included patients undergoing spinal surgery since the introduction of the robotic system (Mazor X Stealth Edition, Medtronic plc). The study evaluated the time to record the patient's anatomy, the fluoroscopy time, radiation dose, implant insertion time, implant accuracy, and complications. RESULTS: The study assessed 67 patients who received 415 pedicle screws (98 thoracic, 284 lumbar, 1 in a transitional vertebra, and 32 sacral). The time to record the patient's anatomy ranged from 4 to 71 (mean: 16.93 ± 12.49) minutes. The fluoroscopy time ranged from 2.3 to 71.20 (mean: 12.91 ± 11.52) seconds. The amount of radiation administered during the surgical procedure ranged from 0.38 to 12.89 (mean: 3.78 ± 3.29) mGy. The time required for implant insertion into the pedicles ranged from 102 to 323 (mean: 147.90 ± 45.09) seconds. In the axial plane, 95.65% of the implants were types A or B on the Gertzbein-Robbins scale, and, in the sagittal plane, 96.38%. No cases required abandonment of the robotic system. As a complication, one patient developed a neurological deficit. CONCLUSION: The robotic system demonstrated high accuracy in implant placement and, with increasing experience in its use, was associated with reductions in fluoroscopy time, intraoperative radiation exposure, and the time to record the patient's anatomy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Robot-assisted lumbar facet joint infiltration improves accuracy and reduces radiation exposure compared to the manual technique in a comparative phantom studyScientific Reports · 2026
- Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study.Journal of robotic surgery · 2026
- Robot-assisted spinal surgery: technological maturity, clinical evidence, and future directions.Frontiers in robotics and AI · 2026
- Robotic-Assisted Spinal Instrumentation from C1 to S1 Experience of a UK Neurosurgical Tertiary Referral Centre.Journal of clinical medicine · 2026
- Axial Convergence in Percutaneous Robotic Pedicle Screw Placement: Mixed-Effects Analysis of 1006 ScrewsSpine Open · 2026
- Robotically equipped 3D hybrid operating room in spine and trauma surgery - updated experience in 359 patients and 2080 screws.Journal of robotic surgery · 2026