Preliminary Study of Oxidation and Reduction for Oxalate Ion Detection

E. Fitriany, Mulyawati, F. Kurniawan

Key points

From the paper's conclusion
  • These findings indicate that modified iron electrodes can detect oxalate ions.

Abstract

The modified iron electrode for detecting oxalate ions was successfully developed. The iron electrode was made by immersing it in a 0.5 M CuSO4 solution. The immersion is carried out using a specific time variation. The iron electrode has been coated and then placed in 50% H2O2 solution to accelerate the oxidation reaction on the electrode surface. The performance of the modified electrode was carried out using the cyclic voltammetry method at a potential range of -0.4 to +0.5 V. Iron electrode without modification was used for comparison study. The result shows no oxidation and reduction peak using an iron electrode without modification, while oxidation and reduction peak appeared at +0.08 V and +0.07 V, respectively, using the modified iron electrode. This data means that the modified iron electrode is sensitive to oxalate ions. The scan rate was varied at 25; 50; 75; 100; and 150 mVs-1. The result showed that the optimum scan rate is 100 mVs -1 . The measurement of 0; 1; 2; 3; 4; 5 mM oxalic ion have also been conducted, showing a linear increase in current (A). The equation is y= 0.0023x +0.0004, R2= 0.9918. The LOD is 6.2 x 10-1 M. This modified electrode is selective toward ascorbic acid, uric acid, and glucose interference. These findings indicate that modified iron electrodes can detect oxalate ions.