From the paper's conclusion- Here, we report a layer-by-layer PAD incorporating controlled interfacial gaps to enhance capillary-driven transport for the determination of dissolved inorganic carbon (DIC) in water.
Abstract
Paper-based analytical devices (PADs) are attractive for point-of-need water analysis because of their low cost and operational simplicity. However, slow fluid transport often limits their analytical throughput. Here, we report a layer-by-layer PAD incorporating controlled interfacial gaps to enhance capillary-driven transport for the determination of dissolved inorganic carbon (DIC) in water. The device employs distance-based colorimetric detection using an acid-base system with bromocresol green (BCG) and was benchmarked against a laminated configuration fabricated from the same Whatman grade 4 paper. The layer-by-layer architecture reduced the colorimetric response time from 18 to 2.5 min while achieving limits of detection and quantification of 0.30 and 0.92 mg L-1, respectively. Analysis of bottled mineral water samples showed good agreement with a reference potentiometric titration method, with relative errors below 7%. These results demonstrate that controlled interfacial engineering can be effectively integrated into distance-based PADs to accelerate analysis without modifying the sensing chemistry, providing a simple and low-cost approach for rapid on-site DIC determination.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Microfluidic Paper-Based Devices at the Edge of Real Samples: Fabrication Limits, Hybrid Detection, and PerspectivesMicromachines · 2026
- A Laminated Microfluidic Paper-Based Analytical Device Based on Image Analysis Using Optimized Color Space for Simultaneous Multiplexed Quantification of Chemical Water Quality Parameters.Analytical chemistry · 2026
- Geometry-driven miniaturization of a microfluidic paper-based analytical device for reagent-efficient colorimetric detection of dopamine.Biomedical microdevices · 2026
- Barcode-based paper analytical device for point-of-care chloride detection with user-independent readout.The Analyst · 2026
- Progress toward real-world diagnostic applications of microfluidic paper-based analytical devices (μPADs)Lab on a Chip · 2026
- A barcode-enabled distance-based paper sensor using silver chloranilate for semi-quantitative determination of chloride in various water sources.Analytical methods : advancing methods and applications · 2026