Lab on a chip

Integration of multi-stacked biodegradable PLGA membranes into microfluidic organ-on-a-chip devices for tunable barrier co-culture.

Minju Kim, Kanghoon Choi, Kibin Park, Laith Al-Rabadi, Jiyoung Chang, Jungkyu Kim

Department of Mechanical Engineering, University of Utah, Salt Lake City, Utah 84112, USA. jkim@mech.utah.edu. · Division of Nephrology, Department of Internal Medicine, University of Utah Health, Salt Lake City, Utah 84112, USA.

Key points

From the paper's conclusion
  • We integrate multi-stacked, thermally fused near-field electrospun PLGA membranes into a microfluidic organ-on-a-chip device.

Abstract

We integrate multi-stacked, thermally fused near-field electrospun PLGA membranes into a microfluidic organ-on-a-chip device. Layer number controls pore diameter (12-33 μm), porosity, circularity and permeability. The biodegradable, optically transparent membranes support endothelial-stromal co-culture with confirmed barrier integrity (TEER, ZO-1) and spatially resolved tissue compartmentalization.