Long-term performance monitoring of a secondary electrospray ionisation high-resolution mass spectrometry platform using a vapor generator for routine quality control.
Albin Vadakkechira, Renato Zenobi, Daniel Pohl, Kathryn J Burton-Pimentel, Stamatios Giannoukos
Department of Chemistry and Applied Biosciences, ETH Zürich, Vladimir-Prelog-Weg 3, Zurich, 8093, Switzerland. stamatios.giannoukos@org.chem.ethz.ch. · Department of Gastroenterology and Hepatology, USZ, Zurich, Switzerland.
Key points
From the paper's conclusion- The implementation of a structured QC protocol provides a valid methodology for long-term SESI-HRMS performance monitoring, ensuring data integrity and supporting the translation of SESI-HRMS into large-scale clinical applications.
Abstract
Secondary electrospray ionisation high-resolution mass spectrometry (SESI-HRMS) is a powerful, non-invasive tool for real-time breath analysis. However, instrumental drift, capillary degradation and other technical variations compromise clinical applications, especially for long-term studies, which require robust quality control (QC) methodologies. In this work, we present the use of an evaporation chamber-based reference gas system to monitor SESI-MS system performance over almost two years. Targeted QC measurements with deuterated butyric acid were recorded three times on each measurement day. Analytical stability was systematically quantified across both positive and negative ion modes. In a standalone continuous characterization, the reference gas generator delivered a stable signal over several hours. A Linear Mixed-Effects Model (LMM) was constructed to evaluate the influence of time-of-day on instrument performance. This revealed significant effects: positive mode signal intensity progressively increased throughout the day, whereas for negative mode, an initial drop in signal intensity and subsequent stabilisation was observed. Furthermore, the longitudinal QC framework successfully tracked capillary performance over time, providing objective performance metrics to retrospectively evaluate the impact of targeted hardware maintenance on system performance. The implementation of a structured QC protocol provides a valid methodology for long-term SESI-HRMS performance monitoring, ensuring data integrity and supporting the translation of SESI-HRMS into large-scale clinical applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Mouthpiece Type Factor: Impact of Filter Selection on Secondary Electrospray Ionisation High-Resolution Mass Spectrometry (SESI-HRMS) Breath Profiles.Journal of breath research · 2026
- Within-day temporal dynamics of exhaled breath features during prolonged fasting monitored by SESI-HRMS: a pilot study.Talanta · 2026
- Secondary Electrospray Ionization Mass Spectrometry for Volatile Analysis: Current Challenges and Emerging SolutionsAnalytica—A Journal of Analytical Chemistry and Chemical Analysis · 2026
- In Vivo Real-Time Metabolic Isotopic Tracing via Breath Analysis upon Deuterated Water Ingestion: A Proof-of-Concept Study.ACS central science · 2026
- Respirator-integrated continuous gas sampling for broadband optical spectroscopy.Scientific reports · 2026
- System Performance Check (SPC): A Readiness Companion to Multi-Attribute Method (MAM) Analysis in Biopharmaceutical Quality Control.Pharmaceuticals (Basel, Switzerland) · 2026