Discrimination transfers but calibration does not across fundus cameras in diabetic retinopathy screening
Sairam Tabibu, Mugdha Abhyankar
Key points
From the paper's conclusion- We asked whether a referable-DR classifier trained on tabletop fundus photographs (BRSET) keeps its calibration — the agreement between predicted probability and observed frequency — as well as its discrimination when applied without adaptation to handheld images (mBRSET) and an independent external dataset (APTOS 2019).
Abstract
Deep-learning screening for referable diabetic retinopathy (DR) increasingly runs on cameras it was not developed on. We asked whether a referable-DR classifier trained on tabletop fundus photographs (BRSET) keeps its calibration — the agreement between predicted probability and observed frequency — as well as its discrimination when applied without adaptation to handheld images (mBRSET) and an independent external dataset (APTOS 2019). Across two networks and five training runs, zero-shot discrimination stayed within 0.03–0.04 AUROC of networks trained on each target (AUROC 0.89–0.96). Calibration did not: on handheld images, expected calibration error (ECE) was 0.06–0.07 without recalibration and 0.10–0.11 with the source-fitted calibrator (0.009 in-domain), and a threshold giving 90% sensitivity on the source detected only 42–47% of referable handheld images. On the handheld target, prevalence matching attributed about four-fifths ( $$79\pm 10\%$$ ) of the calibration loss to its higher disease prevalence; on the external target, prevalence and other dataset differences partly offset each other. Prevalence correction without labels did not restore calibration, whereas Platt or isotonic recalibration with 100–200 labelled target images did (ECE 0.03–0.04). Half-precision models preserved discrimination and calibration; 8-bit integer quantization reduced zero-shot AUROC by 0.07–0.14. New cameras therefore require recalibration on local labels.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Calibrated Confidence and Confidence-Stratified Triage for Diabetic Retinopathy Severity Grading: A Class-Imbalance-Aware Deep Learning Study on the DDR Dataset.Bioengineering (Basel, Switzerland) · 2026
- Calibration failure and limited-label recalibration of retinal image classifiers under dataset shiftDiscover Artificial Intelligence · 2026
- CalibCause: A Quality-Gated, Interpretable and Calibrated Deep Learning Framework for Diabetic Retinopathy GradingarXiv · 2026
- A Modular Diabetic Retinopathy Analysis Framework for Lesion Segmentation and Severity Grading: Cross-Dataset External Validation from APTOS to Messidor-2.Diagnostics (Basel, Switzerland) · 2026
- From Canon to Nikon: Camera Domain Shift and Potential Equity Implications in Retinal AIarXiv · 2026
- Clinical-guided deep learning framework for diabetic retinopathy: integrating lesion-aware attention, adversarial augmentation, and uncertainty quantification.Scientific reports · 2026