Trans-stenotic pressure drop estimation from PC-MRI and ultrasound imaging velocimetry using a modified Bernoulli equation
Ali Amiri, Johan T. Padding, Selene Pirola, Willian Hogendoorn
Delft University of Technology
Key points
From the paper's conclusion- Overall, the results demonstrate that accounting for flow-regime-dependent loss mechanisms enhances pressure drop estimation, and that sufficient sampling of the stenotic throat is crucial for MRI-based flow rate and pressure drop estimation.
Abstract
Abstract Accurate non-invasive estimation of trans-stenotic pressure drops remains a challenge. In clinical practice, pressure drops are often estimated from velocity measurements using Bernoulli-based formulas, but these simplified relations do not explicitly account for how pressure losses change with the flow regime. Here, we introduce a modified Bernoulli (MB) formulation that incorporates regime-dependent pressure losses through a Reynolds-number-dependent loss coefficient. Steady in-vitro experiments were performed in an idealized stenosis model over physiologically relevant flow rates (0.65–3.9 L/min), combining direct pressure measurements with ultrasound imaging velocimetry (UIV) and phase-contrast magnetic resonance imaging (PC-MRI) to measure velocities. The MB model was calibrated from the measured pressure drops and then evaluated against the simplified Bernoulli (SB) and extended Bernoulli (EB) formulations. Over the tested flow regime, MB agreed best with the measurements (typically within about ±10% ). SB and EB showed larger biases, with errors of roughly 10–55% (SB) and -15 to 25%−(EB), and overestimated the pressure drop in the clinically relevant range. We additionally quantified the effect of PC-MRI in-plane pixel size on MRI-based pressure estimates. Coarse in-plane pixel size of 1.33 mm, corresponding to only approximately 3–4 pixels across the 3.3 mm throat radius of the 6.6 mm stenosis throat, led to systematic underestimation of flow rate and bulk velocity (about − 34 to − 44%) and, consequently, of the MB-predicted pressure drop (about − 52 to − 62%). In contrast, the peak throat velocity was substantially less sensitive to pixel size, resulting in smaller estimation errors (about − 13 to − 18.7%) when used as input for the MB. Overall, the results demonstrate that accounting for flow-regime-dependent loss mechanisms enhances pressure drop estimation, and that sufficient sampling of the stenotic throat is crucial for MRI-based flow rate and pressure drop estimation. In addition, peak-velocity-based MB pressure drop estimations are less sensitive to pixel size.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Estimating Pressure Loss Across Stenotic Lesions with Color Doppler.IEEE transactions on ultrasonics · 2026
- Physics-Informed Neural Networks for Stenotic Pressure-Drop Estimation From 4D Flow MRI: Validation Against In Vitro Measurements.International journal for numerical methods in biomedical engineering · 2026
- A Statistical Shape Modeling Approach for the Derivation of a Data‐Driven Geometry‐Aware Lumped Arterial Stenosis ModelInternational Journal for Numerical Methods in Biomedical Engineering · 2026
- Validation of individualized flow simulations for determining the pressure gradient in patients with renal artery stenosismedRxiv · 2026
- Sensitivity of Pulmonary 4D Flow MRI Biomarkers to Controlled Spatial and Temporal Degradation in a CFD BenchmarkbioRxiv · 2026
- Computational assessment of fibromuscular dysplasia-related renal artery stenosisComputers in Biology and Medicine · 2025