Physics-Informed Neural Networks for Stenotic Pressure-Drop Estimation From 4D Flow MRI: Validation Against In Vitro Measurements.
Alexis Caltran, Thomas Puiseux, Barbara Garmy-Susini, Fatima Zohra Mokrane, Ramiro Moreno
Siemens Healthcare SAS, Courbevoie, France. · Spin Up, ALARA Group, Entzheim, France.
Key points
From the paper's conclusion- This study provides a controlled benchmark of PINN‐based pressure reconstruction from 4D Flow MRI‐like data, not a direct clinical validation.
Abstract
Estimating pressure drops non-invasively from 4D Flow MRI velocity data remains challenging: analytical and direct inverse estimators are efficient but sensitive to resolution and noise, whereas Computational Fluid Dynamics (CFD) requires uncertain boundary conditions and substantial preprocessing. Physics-Informed Neural Networks (PINNs) offer a complementary framework by enforcing the Navier-Stokes equations on measured velocity fields, but their robustness under MRI-like data degradation requires validation. We performed an experimental stenosis benchmark to validate a PINN framework for pressure estimation from internal velocity measurements, without prescribed inlet or outlet boundary conditions. Using a 3D-printed stenotic phantom under steady and pulsatile flow conditions, the approach was evaluated against: (1) high-fidelity CFD, (2) synthetic 4D Flow MRI at 1- to 3-mm resolution, and (3) pressure transducer measurements. Against noise-free CFD, the PINN achieved high fidelity (pressure Normalized Root Mean Square Error (NRMSE) < 3 % ). Synthetic MRI tests showed that accuracy was strongly governed by spatial resolution. At 1 mm, pressure-drop errors remained < 5 % for moderate flows but reached ≈ 30 % in high-velocity cases. At 3 mm, degraded flow features led to ≈ 50 % underestimation in steady cases and reduced flow-rate consistency in pulsatile conditions. At adequate resolution, PINN-derived pressure drops agreed with transducer measurements. This study provides a controlled benchmark of PINN-based pressure reconstruction from 4D Flow MRI-like data, not a direct clinical validation. PINNs can recover pressure fields without prescribed boundary conditions, but exhibit resolution-related limitations similar to lightweight pressure estimators such as PPE, STE, and vWERP. Thus, they should be viewed as complementary tools rather than replacements for established estimators. Clinical translation will require resolved imaging and dedicated developments, such as MRI-aware losses or sub-voxel modeling, to mitigate data degradation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Imaging-derived coronary fractional flow reserve: advances in physics-based, machine learning, and physics-informed methods.Physics in medicine and biology · 2026
- Stochastic Representation-Inspired Physics-Informed Neural Networks for Incompressible Fluid FlowsarXiv · 2026
- Cerebrovascular flow field reconstruction method based on residual network of physical information.Frontiers in bioengineering and biotechnology · 2026
- Trans-stenotic pressure drop estimation from PC-MRI and ultrasound imaging velocimetry using a modified Bernoulli equationScientific Reports · 2026
- Estimating Pressure Loss Across Stenotic Lesions with Color Doppler.IEEE transactions on ultrasonics · 2026
- Physics-Informed Neural Networks Simulation and Validation of Airflows in Three-Dimensional Upper Respiratory TractsFluids · 2025