Not Every Semicircle Is Interfacial: High-Frequency Shielding Artifacts in Three-Electrode Electrochemical Impedance Measurements.
Yanjing Zhang, Chenkun Li, Lulu Jia, Xiaoming Li, Fengjiao Yu, Yuping Wu, Jun Huang, Yuhui Chen
State Key Laboratory of Materials-Oriented Chemical Engineering, College of Chemical Engineering, Nanjing Tech University, Nanjing211816, China. · Institute of Energy and Climate Research, IEK-13: Theory and Computation of Energy Materials, Forschungszentrum Jülich GmbH, 52425Jülich, Germany.
Key points
From the paper's conclusion- These results indicate that accurate three-electrode EIS measurements require not only sufficient ionic contact between the reference electrode and electrolyte but also minimal disturbance of the local electric-field distribution.
Abstract
Electrochemical impedance spectroscopy (EIS) is widely used to investigate interfacial processes and transport properties of electrochemical systems. Three-electrode configurations are commonly employed to separate the impedance contributions from individual electrodes; however, the influence of reference electrode structures on high-frequency impedance measurements has received limited attention. In this work, we identify a high-frequency artifact in three-electrode EIS measurements caused by the insulating sheath of the conventional reference electrodes. Using Ag/Ag2O reference electrodes as a model system, an additional high-frequency semicircle was observed in the impedance spectra. Through systematic experiments, this feature is shown to be independent of the electrode materials, electrolyte composition, and instrumental parasitic capacitance. Finite-element simulations further demonstrate that the dielectric sheath around the reference electrode disturbs the local alternating electric-field distribution, resulting in an artificial relaxation process associated with dielectric shielding. The intensity of this artifact depends on the reference electrode geometry and electrolyte dielectric properties, and the same phenomenon is observed in both the aqueous and organic electrolytes. These results indicate that accurate three-electrode EIS measurements require not only sufficient ionic contact between the reference electrode and electrolyte but also minimal disturbance of the local electric-field distribution.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Channel-Anchored Relaxation Maps for Electrochemical Impedance Spectroscopy: A Complementary Perspective to DRT.JACS Au · 2026
- A Guide to Recognizing Your Electrochemical Impedance Spectra: Revisions of the Randles Circuit in (Bio)sensingSensors · 2025
- Electrochemical Impedance Spectroscopy of (Photo)electrochemical Systems: A Practical Guide.ACS measurement science au · 2026
- Comparison of Different Electrochemical Setups for Noninvasive Impedance Measurements on Archeological CoinsChemElectroChem · 2026
- Rapid multichannel electrochemical impedance spectroscopy via sparse-frequency acquisition and embedded spectrum reconstruction.Talanta · 2026
- Analytical–Computational Integration of Equivalent Circuit Modeling, Hybrid Optimization, and Statistical Validation for Electrochemical Impedance SpectroscopyElectrochem · 2025