Talanta

Rapid multichannel electrochemical impedance spectroscopy via sparse-frequency acquisition and embedded spectrum reconstruction.

Zhilong Chen, Zhonghua Yun, Bangyong Sun

School of Intelligence Science and Technology, Xinjiang University, Urumqi 830017, China. · School of Intelligence Science and Technology, Xinjiang University, Urumqi 830017, China. Electronic address: bangyong_sun@xju.edu.cn.

Computational

Key points

From the paper's conclusion
  • These results support rapid multichannel EIS with embedded reference-grid reconstruction under the investigated conditions.

Abstract

Electrochemical impedance spectroscopy (EIS) provides information-rich frequency-dependent measurements but can become a throughput bottleneck when spectra are acquired sequentially across multiple channels. We developed a portable eight-channel EIS platform that combines grouped multichannel acquisition, sparse-frequency measurement, and embedded spectrum reconstruction. Eighteen frequencies (CS18) are physically acquired from a predefined 68-frequency reference grid (FULL68), reducing measured frequency points by 73.53%. RouteC, a lightweight log-frequency PCHIP plus Ridge residual-correction method, reconstructs the corresponding reference-grid complex-impedance spectrum on an STM32H7 microcontroller. The deployed Ridge coefficients were derived from public battery-EIS development data and transferred without aqueous-domain refitting; a controlled 116-pair aqueous-electrochemical dataset was used separately for grouped architecture-level validation. Ridge residual correction reduced complex NRMSE relative to PCHIP-only by 26.3% on the public held-out benchmark and 24.6% in the aqueous grouped analysis. External comparison with a CS3104 workstation showed 1.13% and 0.48% cross-instrument differences for nominal 330 Ω and 1.2 kΩ loads, while the validated 50 Hz-5 kHz RC comparison yielded 3.32% magnitude MAPE, 1.53° phase MAE, and 3.86% complex NRMSE. In four real-water matrices, RouteC-versus-FULL68 complex NRMSE ranged from 0.823% to 2.746%; conductivity-derived TDS-equivalent estimates showed a mean sample-level MAPE of 3.63% relative to a handheld reference. Spectrum-generation time decreased from 75 to 22 s for one channel and from 160 to 48 s for eight channels. These results support rapid multichannel EIS with embedded reference-grid reconstruction under the investigated conditions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Rapid multichannel electrochemical impedance spectroscopy via sparse-frequency acquisition and embedded spectrum reconstruction — Science Skim