Evaluating Limits of Machine Learning-Assisted Raman Spectroscopy in Classification of Biological Samples.
Aman Yadav, Arlin Birkby, Noah Armstrong, Assame Arnob, Ming-Hsun Chou, Alma Fernandez, Aart J Verhoef, Zhenhuan Yi,
Artie McFerrin Department of Chemical Engineering, Texas A&M University, College Station, Texas 77843, United States. · Institute for Quantum Science and Engineering, Texas A&M University, College Station, Texas 77843, United States.
Key points
From the paper's conclusion- Overall, our results demonstrate that data quality and spectral similarity are the primary bottlenecks in ML-assisted Raman spectroscopy.
Abstract
Machine learning (ML)-assisted Raman spectroscopy has become a powerful analytical tool for the classification and identification of analytes; however, technical challenges impacting its detection accuracy have not been thoroughly investigated. This study explores experimental factors affecting classification performance. Among the evaluated ML models, ML algorithms show minimal impact on classification accuracy. Instead, experimental factors, including spectral similarity between tested samples and data quality, dominate detection performance. Increases in spectral noise and spectral similarity significantly reduce classification accuracy. In well-controlled samples with low experimental noise, ML-assisted Raman spectroscopy can discriminate lipid mixtures with a composition difference of 1.85 mol %. To assess the effect of biological heterogeneity, we analyzed single-cell Raman spectra from Saccharomyces cerevisiae strains carrying single, double, or triple gene mutations. Intrinsic cell-to-cell variability introduced substantial spectral differences, severely reducing the accuracy of multiclass classification of these genetically similar strains at the single-cell level. Averaging Raman spectra across multiple cells improved classification accuracy by reducing this spectral variability. We also assess the effectiveness of transfer learning across different Raman spectrometers, specifically by applying an ML model trained on one instrument to another Raman spectrometer. Transfer learning can be improved with proper instrument calibration, highlighting the importance of instrument standardization. Overall, our results demonstrate that data quality and spectral similarity are the primary bottlenecks in ML-assisted Raman spectroscopy. Careful attention to sample preparation, data acquisition, measurement conditions, and instrument calibration is critical to achieving robust and reliable classification performance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Machine Learning-Enhanced Generic Raman Model for Commercial-Scale Cell Culture Process Monitoring.Biotechnology journal · 2026
- BIOM-18 Raman-Based Molecular Fingerprinting of Live Cells Using Machine LearningNeuro-Oncology · 2026
- Label-Free Live Cell Type Prediction by Integrating Raman Spectroscopy and Machine Learning.bioRxiv : the preprint server for biology · 2026
- Wavenumber-Aware Deep Learning for Robust Analysis of Raman SpectraarXiv · 2026
- Variable-specific preprocessing enables cross-scale transferable Raman models from high-throughput bioreactor data.Bioresources and bioprocessing · 2026
- RamEx: an R package for high-throughput microbial ramanome analyses with accurate quality assessmentMicrobiome · 2026