A Practical Approach to Efficiency Calibration of an HPGe Gamma-Ray Spectrometer for Soil Matrix Measurements Using Reference Materials
Deniz Agehan Kahraman, Ayse Nur Esen, Erdi Sirin
Key points
From the paper's conclusion- The results demonstrated that this approach provides a practical, matrix-matched calibration in which geometry-specific effects are inherently accounted for through empirical transfer and verified by reference material validation.
Abstract
In this study, a metrologically traceable and cost-effective efficiency calibration method for a coaxial HPGe gamma-ray spectrometer was developed using the IAEA–312 soil reference material and validated with the IAEA–447 moss-soil reference material. The efficiency curve was derived from selected gamma-ray lines of 214Pb and 214Bi in the energy range 242–1764 keV and fitted using a fourth-order logarithmic polynomial. The fitted model showed excellent agreement with the experimental data (R2 ≈ 0.996), with residuals within ±6%. The calibration was applied to determine the activity concentrations of 226Ra, 232Th, 137Cs, and 40K in IAEA–447. The measured values were in good agreement with the certified values, with relative deviations below 6% and all zeta-scores well below unity, indicating no statistically significant bias and confirming the traceability of the efficiency transfer. A comprehensive uncertainty analysis showed that the uncertainty in the efficiency calibration primarily determines overall accuracy, whereas counting statistics dominate for weaker gamma-ray lines. The results demonstrated that this approach provides a practical, matrix-matched calibration in which geometry-specific effects are inherently accounted for through empirical transfer and verified by reference material validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Monte Carlo uncertainty propagation in efficiency calculations for 4πγ counting of 134 Cs and 137 CsMetrologia · 2026
- Validation of the ray-tracing Monte-Carlo code tRAYcy for geometry and coincidence-summing corrections in environmental gamma-ray spectrometry.Journal of environmental radioactivity · 2026
- Matrix effects on gross alpha determination in environmental water residues: an analytical framework based on effective alpha-particle mass range.Environmental geochemistry and health · 2026
- Testing Efficiency Transfer software in gamma spectrometry: a comparison of different calculation methods for increased complexity of sample geometries.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Optimized Monte Carlo modeling of HPGe detector efficiency with true coincidence summing corrections using PHITS and MCNP5.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Theoretical full-energy peak efficiency determination for LaBr3:Ce detectors.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026