Testing Efficiency Transfer software in gamma spectrometry: a comparison of different calculation methods for increased complexity of sample geometries.

Michele Signoriello, Filippo Bragato, Maria Rosa Fornasier, Sara Savatović, Benedetta Santoro, Alessandro Michele Ferrara, Massimo Giorgio Garavaglia, Mara Severgnini

Department of Medical Physics, Azienda Sanitaria Universitaria Giuliano-Isontina (ASUGI), Via della Pietà 2/2, Trieste, 34129, Italy. Electronic address: michele.signoriello@asugi.sanita.fvg.it. · Regional Agency for Environmental Protection of Friuli Venezia Giulia (ARPA FVG), Via Colunga 42, Udine, 33100, Italy.

Key points

From the paper's conclusion
  • A comprehensive evaluation of the most used Efficiency Transfer software was conducted to assess their performance in gamma-ray spectrometry. This work aims to enhance the accuracy of activity concentration estimates, for samples with non-standard geometries, thus enabling rapid and reliable responses in potential radiological emergencies.

Abstract

BACKGROUND: Gamma spectrometry with high-resolution High-Purity Germanium detectors is a widely adopted technique, valued for its excellent energy resolution. It is applied across a range of Radiation Protection activities, including the measurement of low-activity concentrations in environmental samples containing natural and artificial radionuclides, as well as quality assurance assessments of radiopharmaceuticals used in diagnostic imaging. METHODS: This work presents an inter-comparison between software for Efficiency Transfer calculations, including three semi-empirical (EFFTRAN, ETNA, and ANGLE) and two Monte Carlo based codes (GESPECOR and LabSOCS/ISOCS). The calculations considered sample geometries of increasing complexity, ranging from a Marinelli beaker to a vial and wipe test. A verification test (IAEA-TERC-2022-01/02) was conducted to assess the accuracy of the simulations by validating the results against a Verification Sample with well-known low activity radionuclide concentrations. RESULTS: Calculations performed for the different sample geometries yielded consistent Efficiency Transfer coefficients across all the software tested. The largest relative bias was observed at a low photon energy (59keV from 241Am), which exceeded the 20% acceptability limit for the more complex sample under investigation. A similar trend was observed in the activity concentration results obtained from the verification test. CONCLUSIONS: A comprehensive evaluation of the most used Efficiency Transfer software was conducted to assess their performance in gamma-ray spectrometry. This work aims to enhance the accuracy of activity concentration estimates, for samples with non-standard geometries, thus enabling rapid and reliable responses in potential radiological emergencies.