Traceable Measurement of Peak Irradiance and Fluence Rate Enables Accurate Dose Determination of Pulsed Xenon Ultraviolet Disinfection.
Mengkai Li, Jiale Wang, Yanyan Huang, Zhimin Qiang, Guoliang Li, Zhe Sun, Yue Wang, Yanfei Wang,
State Environmental Protection Key Laboratory of Environmental Health Impact Assessment of Emerging Contaminants School of Environmental Science and Engineering, Shanghai Jiao Tong University, Shanghai200240, China. · Key Laboratory of Environmental Aquatic Chemistry, State Key Laboratory of Regional Environment and Sustainability, Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences, Beijing100085, China.
Key points
From the paper's conclusion- We established dose-response curves for Escherichia coli and bacteriophage MS2 and quantified pulse requirements for 2-log10 inactivation.
Abstract
Pulsed xenon ultraviolet (PX-UV) disinfection delivers UV energy as microsecond-scale pulses with ultrahigh peak irradiation, creating a need for pulse-resolved dose characterization. Under such conditions, conventional UV dose formulation obscures peak exposure and the pulse temporal structure within an integrated dose, while commonly used radiometers and chemical actinometers are limited by saturation, nonlinear response, spectral mismatch, and temporal averaging. In this study, we addressed these limitations by developing a radiation-hard cerium-doped fiber sensor (Ce-FS) and combining it with broadband chemical actinometric calibration to enable traceable measurements of peak irradiance (IR0, for surface disinfection) and fluence rate (FR0, for air disinfection) under PX-UV irradiation. On this basis, we reformulated the dose to explicitly incorporate IR0 or FR0, pulse duration, and frequency. The Ce-FS showed a wide linear range (3.90-4611.66 mW cm-2), was benchmarked against conventional sensors, and showed good agreement with theoretical predictions. We established dose-response curves for Escherichia coli and bacteriophage MS2 and quantified pulse requirements for 2-log10 inactivation. The kinetic transferability of the pulse-resolved dose was demonstrated through a dose-controlled disinfection trial in a hospital intensive care unit. These results provide a solid basis for rigorous dose determination in PX-UV disinfection, thereby supporting performance evaluation, standardization, and dose-controlled disinfection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Monochromatic Short-Wave UV Sensitivity of Pathogenic Bacteria.Bulletin of experimental biology and medicine · 2026
- UV-C disinfection of endocavitary ultrasound probes against Candidozyma auris : a rapid and chemical-free solutionAntimicrobial Stewardship & Healthcare Epidemiology · 2026
- UVC-LED Disinfection Technology: Mainstream Waveband Characteristics, Applications and ChallengesWater Air & Soil Pollution · 2026
- Defining Operational UV-C Dose Requirements for Autonomous Disinfection of Clinically Relevant Pathogens Across Healthcare and High-Touch SurfacesbioRxiv · 2026
- Radiometrically Equivalent Comparison of 254 nm Low-Pressure Mercury Lamps and InGaN 280 nm UV-C LEDs for Surface Disinfection.ACS omega · 2026
- X-ray inactivation of SARS-CoV-2 for diagnostic testing workflows in pandemic responseScientific Reports · 2026