Calibration-Free Analysis of Li Isotope Ratios Using Laser Ablation and Laser Absorption Spectroscopy

Mark C. Phillips, Kyle A. Makovsky, R. E. Stevens, Elizabeth J. Kautz, S. S. Harilal

University of Arizona · Pacific Northwest National Laboratory

Key points

From the paper's conclusion
  • Isotope ratios determined from the spectral fits show relative accuracy within -0.3% to -5% of ICP-MS measurements, without requiring calibration to external standards, and offering a rapid analysis approach to lithium isotope determination in solid materials relevant to nuclear energy, safeguards, and geochemistry.

Abstract

Li ratios demonstrates relative isotopic precisions of 0.6-1.8% for spectra acquired in 30 s. Isotope ratios determined from the spectral fits show relative accuracy within -0.3% to -5% of ICP-MS measurements, without requiring calibration to external standards, and offering a rapid analysis approach to lithium isotope determination in solid materials relevant to nuclear energy, safeguards, and geochemistry.