Multicenter evaluation of total body irradiation planning strategies: A DGMP working group study on large-field irradiation techniques.
Maya Shariff, Heiko Karle, Oliver Schramm, Johann Brand, Andre Karius, Sven Knobe, Hendrik Auerbach, Jerome Krayenbuehl,
Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany; Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Erlangen, Germany. Electronic address: maya.shariff@uk-erlangen.de. · Department of Radiation Oncology, University Medical Center Mainz, Mainz, Germany.
Key points
From the paper's conclusion- This study provides the first standardized multicenter planning-based comparison of contemporary TBI approaches in Germany and Switzerland. By quantifying inter-institutional variability in dose distributions under uniform conditions, the results outline the current heterogeneity of clinical practice and the range of achievable dosimetric outcomes. Rather than defining a reference standard, these findings establish a structured benchmark of existing planning strategies and highlight the need for harmonization, thereby providing a foundation for future consensus development and end-to-end validation efforts.
Abstract
PURPOSE: Total body irradiation (TBI) is an integral component of conditioning regimens for hematopoietic stem cell transplantation. Despite similar biological objectives, planning and delivery of TBI vary substantially across institutions. The aim of this study was to quantitatively assess inter-institutional and technique-dependent variability in contemporary TBI planning by comparing dose distributions generated under standardized conditions across multiple centers in Germany and Switzerland. METHODS: Twenty-four radiotherapy centers received anonymized total-body computed tomography (CT) datasets of a standardized dummy patient in head-first supine and prone orientations, including predefined points of interest (POI) and organs at risk (OARs). Each center was asked to generate TBI plans for two prescription levels (8 Gy in 4 fractions and 12 Gy in 6 fractions) using its local planning standards, with the entire body defined as the planning target volume (PTV). Submissions included treatment-planning-system (TPS)-based 3D dose distributions and in-house table-based calculations. TPS plans were centrally imported into a TPS for extraction of target coverage and OAR dose metrics, while table-based submissions were evaluated using aggregated POI doses. RESULTS: In total, 45 TBI treatment plans from 24 centers were analyzed, including 24 plans for 8 Gy and 21 plans for 12 Gy. Both TPS-based and table-based planning approaches were represented across a broad range of clinical techniques, including volumetic modulated arc therapy (VMAT), Tomotherapy, sweeping beam, static field, bilateral, and translation couch setups. Marked inter-institutional variability was observed, particularly for lung, lens, and kidney doses. At the 12 Gy prescription level, mean lung doses differed by more than 3 Gy between centers, corresponding to relative variations of up to approximately 30%. Conformal intensity-modulated techniques (VMAT and Tomotherapy) showed the highest consistency across institutions, whereas non-conformal approaches exhibited substantially larger dose variability. Increased lung sparing at 12 Gy was consistently associated with reduced target coverage compared to 8 Gy. CONCLUSION: This study provides the first standardized multicenter planning-based comparison of contemporary TBI approaches in Germany and Switzerland. By quantifying inter-institutional variability in dose distributions under uniform conditions, the results outline the current heterogeneity of clinical practice and the range of achievable dosimetric outcomes. Rather than defining a reference standard, these findings establish a structured benchmark of existing planning strategies and highlight the need for harmonization, thereby providing a foundation for future consensus development and end-to-end validation efforts.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Total body irradiation practice patterns: insights from a national survey.Frontiers in oncology · 2026
- End-to-end technical validation of multi-isocenter total body irradiation using an anthropomorphic phantom.Radiological physics and technology · 2026
- Fictive Field-based Feathering for Extended-SSD Total Body Irradiation: A Single-institution Planning and Dosimetric Analysis.Journal of medical physics · 2026
- Total body irradiation across conditioning intensities: historical context, dose-fractionation strategy, and the role of the chemotherapy backbone.Bone marrow transplantation · 2026
- Feasibility, Treatment Planning, and Clinical Results of VMAT-Based Total Body Irradiation for Pediatric Patients: Single-Center Experience.Pediatric blood & cancer · 2026
- Quantifying Patient Setup Error in Total Body Irradiation With Megavoltage Computed Tomography-guided Helical Tomotherapy.In vivo (Athens, Greece) · 2026