Transcranioplasty Ultrasonography as a Point-of-Care Imaging Tool for Residual Subdural Hematoma Surveillance: A Preliminary Validation Study.

Mark Constable, Joshua A Cuoco, Nicholas Tomson, Laura Milosevic, Nathan Ritchey, Victoria Schunemann, Shahid M Nimjee, Patrick P Youssef,

Department of Neurological Surgery, The Ohio State University Wexner Medical Center, Columbus, Ohio, USA. Electronic address: mark.constable@osumc.edu. · Department of Neurological Surgery, The Ohio State University Wexner Medical Center, Columbus, Ohio, USA.

Observational · n=16

Key points

From the paper's conclusion
  • In the present study, we found that TCUS provides highly correlated and reliable measurements of residual SDH as compared to CT imaging. Large-scale, prospective studies are necessary to validate these preliminary findings.

Abstract

BACKGROUND: Currently, there is no point-of-care tool to monitor residual subdural hematoma (SDH) following surgical evacuation. Transcranioplasty ultrasonography (TCUS) via a sonolucent burr-hole cover can mitigate the limitations of traditional transcranial ultrasound; however, the reliability of TCUS, compared with CT, in measuring SDH remains unknown. Here, we compare residual SDH measurements following surgical evacuation between TCUS and CT. METHODS: We conducted a retrospective study of 16 consecutive patients with 19 unique SDHs. Comparative measurements of residual SDH following surgical evacuation were performed between TCUS and CT. A total of 40 unique comparisons were performed throughout the follow-up period. The distance from plate surface to anatomic midline and distance from plate surface to cortical surface (i.e., residual SDH) were compared between modalities. Pearson correlation coefficients, Lin's concordance correlation coefficients, and Bland-Altman plots were calculated. RESULTS: The average distance to midline measured on TCUS was 63mm compared to 64mm on CT, with a positive linear correlation between techniques (r = 0.966, P < 0.001). The average distance to cortex as measured on TCUS was 14mm compared to 14mm on CT, with a positive linear correlation between modalities (r = 0.966, P < 0.001). Lin's concordance correlation coefficient was 0.948 (95% CI: 0.904-0.972) and 0.961 (95% CI: 0.928-0.979) between techniques for distance to midline and cortex, respectively. CONCLUSIONS: In the present study, we found that TCUS provides highly correlated and reliable measurements of residual SDH as compared to CT imaging. Large-scale, prospective studies are necessary to validate these preliminary findings.

Read the paper