A Novel Technique for Non-Invasive Measurement of the Cerebral Metabolic Rate of Oxygen in the Rat.

Tonja W Emans, Mandy Zhang, Khushi Kothari, Debra Fong, Carolyn J Barrett, David J Dubowitz, Fiona D McBryde

Cerebrovascular Control Laboratory, Pūtahi Manawa Centre of Research Excellence, Department of Physiology, University of Auckland, Auckland, New Zealand. · Centre for Advanced Magnetic Resonance Imaging, Faculty of Medical and Health Sciences, University of Auckland, New Zealand.

Preclinical

Key points

From the paper's conclusion
  • Our results confirm that our non-invasive measures of SaO2 and SvO2 and the resulting CMRO2 values are comparable with the results from the current gold standard methods. Thus, our approach using PAI and pulse oximetry to measure CMRO2 appears to be viable.

Abstract

OBJECTIVE: The brain is dependent on oxidative metabolism. Changes in the cerebral metabolic rate of oxygen (CMRO2) are implicated in many neurologic diseases but generally require invasive measurement. We aimed to validate a novel method to non-invasively assess CMRO2 in rats using pulse oximetry and photoacoustic imaging (PAI) to determine the arteriovenous oxygen difference across the brain. METHODS: Under anaesthesia, Wistar rats were instrumented to record the internal carotid blood flow (iCBF) and cannulated for blood sampling. Arterial oxygen saturation (SaO2) and venous oxygen saturation (SvO2) were measured at different fractions of inspired oxygen (FiO2: 100%-21%), via pulse oximetry and PAI of the superior sagittal sinus vein, using the Vevo LAZR-X. The CMRO2 was calculated using arteriovenous oxygen differences and the iCBF. Results were validated against blood gas co-oximetry. RESULTS: Bland-Altman analyses revealed a high degree of agreement between co-oximetry and PAI/pulse oximetry. A significant decrease in the SvO2 and SaO2 at an FiO2 of 21% was detected by both methods (p < 0.0001) and the resultant CMRO2 values were comparable amongst the techniques. However, we found some underestimation of SvO2 with PAI versus co-oximetry at an FiO2 of 70% and 100%. Induced hypothermia was used to validate the use of non-invasive PAI/oximetry combined with the iCBF by ultrasound imaging and showed a CMRO2 decrease from 6.4 ± 3.0 to 4.1 ± 1.9 mL O2/100 g/min (p = 0.0114). CONCLUSION: Our results confirm that our non-invasive measures of SaO2 and SvO2 and the resulting CMRO2 values are comparable with the results from the current gold standard methods. Thus, our approach using PAI and pulse oximetry to measure CMRO2 appears to be viable.

Read the paper