A wearable ultrasonic system for continuous, non-invasive blood pressure monitoring and knowledge-based hemodynamic analytics.

Yu Chu, Huaiyu Wu, Xiangming Xue, Edward Chan, Nitin Sharma, Ruihai Zhou, Xiaoning Jiang

Department of Mechanical and Aerospace Engineering, North Carolina State University, Raleigh, NC 27695, USA. · Department of Mechanical and Aerospace Engineering, North Carolina State University, Raleigh, NC 27695, USA; Joint Department of Biomedical Engineering, North Carolina State University, Raleigh, NC 27695, USA.

Key points

From the paper's conclusion
  • These findings support combining compact single-element sensing with reference-based pressure reconstruction and knowledge-based waveform analysis to characterize changes within individuals.

Abstract

Continuous blood-pressure (BP) waveforms reveal pressure changes between heartbeats that intermittent cuff readings miss. A compact wearable ultrasonic system was developed to reconstruct these waveforms from superficial arterial wall motion and analyze their features. A 6.5-MHz single-element transducer recorded repeated radio-frequency (RF) echoes over the ulnar artery. An exponential pressure-area model converted the separation of two candidate wall echoes to pressure. The reconstructed waveform then supplied the input for feature analysis, with knowledge-based rules checking signal quality, feature availability, timing and calibration before selecting descriptors for reporting. In 33 resting and seated post-activity recordings from three participants, pressure was reconstructed with a fixed mapping for each participant. Resting systolic BP had between-record standard deviations of 0.7-1.9 mmHg. In the first seated recordings after squats and running, systolic BP rose by 15.3-29.2 mmHg and pulse rate by 20.2-33.7 bpm relative to the corresponding pre-activity baselines. Recovery varied among participants. Both increases persisted across the tested analysis windows and peak-detection settings. A separate six-session dataset from Participant 01 retained 36 analysis windows after quality screening, supporting reconstruction after repeated placements. Analysis of an accepted waveform retained systolic, diastolic and interval descriptors while excluding inconsistent notch features. These findings support combining compact single-element sensing with reference-based pressure reconstruction and knowledge-based waveform analysis to characterize changes within individuals.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers