Analytical agreement of a multiparametric, fully quantitative fluorescent immunoassay analyser (Vcheck) with commercial laboratory assays for NT‑proBNP and cardiac troponin‑I in dogs and cats.

Luca Ferasin, Emma Hudson, Jo Farminer, Heidi Ferasin

Specialist Veterinary Cardiology Consultancy, Hampshire, United Kingdom. Electronic address: luca@cardiospecialist.co.uk. · The Ralph Veterinary Referral Centre, Buckinghamshire, United Kingdom.

Key points

From the paper's conclusion
  • Vcheck demonstrates analytical agreement with reference laboratory assays for NT-proBNP and cTnI, with substantial advantages in turnaround time and lower per-test consumable costs. These findings support its integration into small-animal cardiology practice, particularly where rapid decision-making is essential.

Abstract

BACKGROUND: Cardiac biomarkers, such as N-terminal pro-B-type natriuretic peptide (NT-proBNP) and cardiac troponin-I (cTnI), are widely used in dogs and cats to assess myocardial stretch and injury, respectively. However, most measurements rely on external laboratories, resulting in turnaround times of 2-7 days. The Vcheck analyser is a fully quantitative, automated fluorescent immunoassay platform capable of providing rapid point-of-care results, but independent validation using robust statistical methodology is required. OBJECTIVES: To evaluate the analytical agreement between Vcheck NT-proBNP and cTnI measurements and established commercial assays (Cardiopet® and Access®), and to assess clinical classification concordance, turnaround time, and per-test consumable cost comparison. METHODS: Paired serum samples from 60 cats and 60 dogs were analysed using Vcheck and corresponding commercial assays. Agreement was assessed using Bland-Altman analysis. Out-of-range values were excluded from Bland-Altman analysis but retained for clinical classification. RESULTS: NT-proBNP demonstrated minimal proportional differences and excellent clinical classification agreement in both species (cats: 100% across all categories; dogs: 88.2-96.0%). Cardiac TnI showed proportional bias, but classification agreement remained high, particularly for elevated and high values (cats: 78.8-100%; dogs: 70.0-94.7%). Vcheck provided results within 10-15minutes, compared with 2-7 days for commercial assays, and per-test costs were substantially lower. CONCLUSIONS: Vcheck demonstrates analytical agreement with reference laboratory assays for NT-proBNP and cTnI, with substantial advantages in turnaround time and lower per-test consumable costs. These findings support its integration into small-animal cardiology practice, particularly where rapid decision-making is essential.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers