Analytical agreement of a multiparametric, fully quantitative fluorescent immunoassay analyser (Vcheck) with commercial laboratory assays for NT‑proBNP and cardiac troponin‑I in dogs and cats.
Luca Ferasin, Emma Hudson, Jo Farminer, Heidi Ferasin
Specialist Veterinary Cardiology Consultancy, Hampshire, United Kingdom. Electronic address: luca@cardiospecialist.co.uk. · The Ralph Veterinary Referral Centre, Buckinghamshire, United Kingdom.
Key points
From the paper's conclusion- Vcheck demonstrates analytical agreement with reference laboratory assays for NT-proBNP and cTnI, with substantial advantages in turnaround time and lower per-test consumable costs. These findings support its integration into small-animal cardiology practice, particularly where rapid decision-making is essential.
Abstract
BACKGROUND: Cardiac biomarkers, such as N-terminal pro-B-type natriuretic peptide (NT-proBNP) and cardiac troponin-I (cTnI), are widely used in dogs and cats to assess myocardial stretch and injury, respectively. However, most measurements rely on external laboratories, resulting in turnaround times of 2-7 days. The Vcheck analyser is a fully quantitative, automated fluorescent immunoassay platform capable of providing rapid point-of-care results, but independent validation using robust statistical methodology is required. OBJECTIVES: To evaluate the analytical agreement between Vcheck NT-proBNP and cTnI measurements and established commercial assays (Cardiopet® and Access®), and to assess clinical classification concordance, turnaround time, and per-test consumable cost comparison. METHODS: Paired serum samples from 60 cats and 60 dogs were analysed using Vcheck and corresponding commercial assays. Agreement was assessed using Bland-Altman analysis. Out-of-range values were excluded from Bland-Altman analysis but retained for clinical classification. RESULTS: NT-proBNP demonstrated minimal proportional differences and excellent clinical classification agreement in both species (cats: 100% across all categories; dogs: 88.2-96.0%). Cardiac TnI showed proportional bias, but classification agreement remained high, particularly for elevated and high values (cats: 78.8-100%; dogs: 70.0-94.7%). Vcheck provided results within 10-15minutes, compared with 2-7 days for commercial assays, and per-test costs were substantially lower. CONCLUSIONS: Vcheck demonstrates analytical agreement with reference laboratory assays for NT-proBNP and cTnI, with substantial advantages in turnaround time and lower per-test consumable costs. These findings support its integration into small-animal cardiology practice, particularly where rapid decision-making is essential.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic performance of Vcheck NT-proBNP and cardiac troponin I assays for detection of dilated cardiomyopathy in Doberman Pinschers.Frontiers in veterinary science · 2026
- Point-of-care troponin I assay has acceptable performance detecting clinically important canine myocardial injury but reduced precision at low concentrations.American journal of veterinary research · 2026
- Analytical validation and clinical evaluation of a high-sensitivity cardiac troponin I immunoassay for use in catsPLoS ONE · 2026
- Evaluation of a Commercially Available Canine C-Reactive Protein Assay Across Institutions and After 2 Years of Frozen Storage.Veterinary clinical pathology · 2026
- Clinical triage concordance of a point-of-care high-sensitivity cardiac troponin I assay using the ESC 0/1-hour algorithm.Clinica chimica acta; international journal of clinical chemistry · 2026
- Method Comparison of PT, INR, and aPTT Results Obtained Using Two Coagulation Analyzers in Routine Laboratory Practice.Diagnostics (Basel, Switzerland) · 2026