A Digital Ultrasonic Height Measurement Device for Children Aged Under Five Years show high reliability yet variable accuracy.
Ms Shan Huang, Prof Caroline Se Homer, Dr Joel Conkle, Dr Sayaka Horiuchi, Dr Damian Pavlyshyn, Prof Joshua P Vogel
Monash University Public Health and Preventative Medicine 553 St Kilda Road Melbourne, Australia; Women's, Children's and Adolescent's Health Program, Burnet Institute 85 Commercial Rd, Melbourne, Australia. Electronic address: shan.huang@monash.edu. · Monash University Public Health and Preventative Medicine 553 St Kilda Road Melbourne, Australia; Women's, Children's and Adolescent's Health Program, Burnet Institute 85 Commercial Rd, Melbourne, Australia.
Key points
From the paper's conclusion- We found that the ultrasonic device showed high reliability for children under 5 years when compared to the board, though accuracy was variable, with greater bias found in older children. Further testing is required to identify the source of inaccuracies and also evaluate reliability in other contexts.
Abstract
BACKGROUND: Accurate height measurement is essential for monitoring child growth. OBJECTIVE: Our aim was to assess the reliability and accuracy of a handheld ultrasonic measurement device compared with a standard measurement board METHODS: We used a cross-sectional methods-comparison study design with experts measuring n=57 children aged 0-5 years in metropolitan Melbourne, Australia. The ultrasonic device was our index test and the measurement board, our reference test. RESULTS: The ultrasonic device demonstrated a similar reliability to the board with a technical error of measurement (TEM) of 0.25cm (95%CI: 0.22, 0.28cm) and 0.19cm (95%CI: 0.17, 0.22) respectively, both within current standards. F-test ratios (ultrasonic vs. board) for intra-measurer TEMs were higher for children ≤2 years at 1.39 (95% CI: 1.03, 1.87; p=0.03) than children >2 years at 1.20 (95%CI: 0.94, 1.51; p=0.14). The ultrasonic device measured children slightly taller than the board (mean difference 0.20 cm; 95% CI: 0.02, 0.38; SD=0.67; p<0.01), with a larger difference in older children who were measured standing (0.35 cm; p=0.01). The Bland Altman 95% limits of agreement ranged from -1.13cm to 1.53cm, demonstrating variance in the measurements between the different tools. CONCLUSION: We found that the ultrasonic device showed high reliability for children under 5 years when compared to the board, though accuracy was variable, with greater bias found in older children. Further testing is required to identify the source of inaccuracies and also evaluate reliability in other contexts.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Non‐Soundproof Tablet Audiometry in Pediatrics: Comparison With Traditional AudiometryThe Laryngoscope · 2026
- The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System.Diagnostics (Basel, Switzerland) · 2026
- Comparing Subglottic Ultrasound and Distal Phalanx Width for Pediatric Endotracheal Tube Sizing in a Predominantly School-Aged Pediatric Cohort: A Prospective Study.Paediatric anaesthesia · 2026
- Height estimation in intensive care: A comparative study.Australian critical care : official journal of the Confederation of Australian Critical Care Nurses · 2026
- A Multicenter Study for Validation of the New Pediatric Advanced Life Support Tape for Indian Children.Indian pediatrics · 2026
- Accuracy Evaluation of the Omron HEM-7120 Automated Oscillometric Upper-arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060-2:2018 Numerical Criteria.Journal of clinical hypertension (Greenwich, Conn.) · 2026