A Digital Ultrasonic Height Measurement Device for Children Aged Under Five Years show high reliability yet variable accuracy.

Ms Shan Huang, Prof Caroline Se Homer, Dr Joel Conkle, Dr Sayaka Horiuchi, Dr Damian Pavlyshyn, Prof Joshua P Vogel

Monash University Public Health and Preventative Medicine 553 St Kilda Road Melbourne, Australia; Women's, Children's and Adolescent's Health Program, Burnet Institute 85 Commercial Rd, Melbourne, Australia. Electronic address: shan.huang@monash.edu. · Monash University Public Health and Preventative Medicine 553 St Kilda Road Melbourne, Australia; Women's, Children's and Adolescent's Health Program, Burnet Institute 85 Commercial Rd, Melbourne, Australia.

Observational · n=57

Key points

From the paper's conclusion
  • We found that the ultrasonic device showed high reliability for children under 5 years when compared to the board, though accuracy was variable, with greater bias found in older children. Further testing is required to identify the source of inaccuracies and also evaluate reliability in other contexts.

Abstract

BACKGROUND: Accurate height measurement is essential for monitoring child growth. OBJECTIVE: Our aim was to assess the reliability and accuracy of a handheld ultrasonic measurement device compared with a standard measurement board METHODS: We used a cross-sectional methods-comparison study design with experts measuring n=57 children aged 0-5 years in metropolitan Melbourne, Australia. The ultrasonic device was our index test and the measurement board, our reference test. RESULTS: The ultrasonic device demonstrated a similar reliability to the board with a technical error of measurement (TEM) of 0.25cm (95%CI: 0.22, 0.28cm) and 0.19cm (95%CI: 0.17, 0.22) respectively, both within current standards. F-test ratios (ultrasonic vs. board) for intra-measurer TEMs were higher for children ≤2 years at 1.39 (95% CI: 1.03, 1.87; p=0.03) than children >2 years at 1.20 (95%CI: 0.94, 1.51; p=0.14). The ultrasonic device measured children slightly taller than the board (mean difference 0.20 cm; 95% CI: 0.02, 0.38; SD=0.67; p<0.01), with a larger difference in older children who were measured standing (0.35 cm; p=0.01). The Bland Altman 95% limits of agreement ranged from -1.13cm to 1.53cm, demonstrating variance in the measurements between the different tools. CONCLUSION: We found that the ultrasonic device showed high reliability for children under 5 years when compared to the board, though accuracy was variable, with greater bias found in older children. Further testing is required to identify the source of inaccuracies and also evaluate reliability in other contexts.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers