Optical crosstalk between SiPMs in dual-ended readout.
Nicolaus Kratochwil, Emilie Roncali, Joshua W Cates, Gerard Ariño-Estrada
Department of Biomedical Engineering, University of California, Davis, Davis, 95616, CA, United States of America. · Lawrence Berkeley National Laboratory, Berkeley, 94720, CA, United States of America.
Key points
From the paper's conclusion- We studied the impact of external optical crosstalk in a dual-ended readout setup using AFBR-S4N44P014M silicon photomultipliers (SiPMs) from Broadcom coupled to 20 mm thick bismuth germanate (BGO) crystal, with and without scintillation light.
Abstract
Gamma-ray detection performance in scintillation crystals can be improved by coupling multiple photodetectors around the sensitive volume. Greater light collection leads to improved time and energy resolution. However, neighboring detector noise, both uncorrelated and correlated, can produce optical photons that may deteriorate their performance. We studied the impact of external optical crosstalk in a dual-ended readout setup using AFBR-S4N44P014M silicon photomultipliers (SiPMs) from Broadcom coupled to 20 mm thick bismuth germanate (BGO) crystal, with and without scintillation light. SiPMs were biased individually. The behavior of crosstalk was analytically modeled. The dark count rate increased up to 50 % when both SiPMs were biased with respect to only one active SiPM. At the same time, the crosstalk probability increased more than 2-fold at high overvoltages due to detection and re-emission of avalanche photons, limiting the operational range and causing an up to 5-fold increase in the total number of detected photons. Energy resolution in BGO was impacted, and the coincidence time resolution distribution shape between two BGO detectors was significantly altered- both in a non-intuitively manner. Experimental results validate the analytical model which rely on a single constant to predict dark count rate, crosstalk and light output enhancement. Crosstalk in multi-SiPM configurations, especially at high overvoltage, significantly impacts performance. Deep understanding of the effects of excess noise in SiPM is crucial for an optimal overall gamma detector operation. Findings extend to applications in high-energy physics, time-of-flight positron emission tomography using dual-ended or monolithic crystals, and cryogenic SiPM experiments in neutrino/dark matter studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Neutron irradiation damage on Silicon Photomultipliers and in-situ current annealing studies for the CBM RICH detectorNuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 2026
- SiPM optical readout of a PICOSEC Micromegas precise timing detectorarXiv · 2026
- Optical readout of MicroPattern Gaseous Detectors with SiPMsarXiv · 2026
- A dual-material structured light guide for improving flood map quality in pixelated LYSO-SiPM detectorsJournal of Instrumentation · 2026
- High-energy gamma-ray spectrometry with LYSO:Ce scintillators for fusion plasma experiments.The Review of scientific instruments · 2026
- High spatial resolution dual-ended readout PET detector based on continuous resistive layer silicon photomultiplier.Physics in medicine and biology · 2026